16
knob at a desired value, the tester will do the test continuously. During the test, if the test
current is great than the preset value, the tester will alarm with sound and light. Set the
“RESET”signal enable, the alarm will be stopped. Replace the DUT and continue to test.
After testing, enable the “RESET” signal to stop the test.
(
2
)
Automatic test
If the test time is not set to 0, the tester will enter into the automatic test. After the PLC
“TEST” signal enabled, the test light will be lit; adjust the output voltage knob at a desired
value, the tester will do the test. During the test, if the test current is great than the preset
value, the tester will alarm with sound and light. Set the PLC “RESET” signal enabled, the
alarm will be stopped. If the test time is return to 0 and the tester is not alarm, the test is
passed and the DUT (Device Under Test)T is eligible.
Note: During DCW test, confirm the measurement circuit is without any electricity before
replace another device.
IX.Precautions
9.2 Maintenance
9.2.1Warranty
1.
Maintenance period
:
The instrument is warranted to be free from
use for a period of 12 months from the date of shipment to the original
end users in different sales spots.
2.
Accessories warranty period is six month.
9.2.2 Maintenance
Please bring forth the warranty card while maintaining. The company
provides lift-long maintenance service to all the shipped instruments. In
this period, consumer is responsible for the maintaining fee if the
instrument is damaged by improper operation.
ATTENTION FOR SAFTY
9.1
During the test, the operator can not touch the position or area described below;
otherwise it will cause electric shock accident.
(
1
)
High-voltage output port of the tester
;