HT3302
EN - 9
Fig. 9: Laser pointer activation
12. The symbol “
” flashes on the display. Move the
J
key to “+” or “-” to enable (ON) or
disable (OFF) the laser pointer upon pressing the
T
key.
13. Press the
J
key to confirm selection and go on to next item.
Fig. 10: MAX, MIN, MAX-MIN, AVG function activation
14. The symbol “MAX-MIN AVG” flashes on the display. Move the
J
key to “+” or “-” to
enable (ON) or disable (OFF) the reading function of MAX, MIN, AVG and MAX-MIN
value on the display during and at the end of measurement (see § 5.2).
15. Press the
J
key to confirm selection and go on to next item.
Fig. 11: HIGH and LOW alarm function activation
16. The symbol “Alarm HI” flashes on the display. Move the
J
key to “+” or “-” to set the
value of the maximum alarm threshold on temperature measurement between values
-
30°C
and
650°C
or “
OFF
” (disabled). Keep the
J
key pressed in position “+” or “-” to
quickly scroll through the values. The buzzer will sound continuously if the measured
temperature value is higher than the HIGH threshold.
17. Press the
J
key to confirm the HIGH threshold setting and go on to the LOW threshold
(see Fig. 11 – right side).
18. The symbol “Alarm LO” flashes on the display. Move the
J
key to “+” or “-” to set the
value of the minimum alarm threshold on temperature measurement between values
-
30°C
and
650°C
or “
OFF
” (disabled). Keep the
J
key pressed in position “+” or “-” to
quickly scroll through the values. The buzzer will sound continuously if the measured
temperature value is lower than the LOW threshold.
19. Press the
J
key to confirm selection and go on to next item.