T2000-T2100
EN – 10
5.2.2. Clamp operation check
1. Press the
ON/OFF
key to switch on the instrument
2. The displayed message “
OL
” indicates that the instrument is ready to carry out
measurements
3. Open the jaws gently (the display will show the screen in Fig. 7) and clamp one test loop
provided as accessory (see Fig. 8).
Fig. 7
Fig. 8: Resistance measurement of test loop
4. Check that the test resistance value is equal to
5.0
(for test loop of 5
). A value
measured by the instrument showing a difference of
0.3
with respect to the rated
value is acceptable (a display of 4.7
or 5.3
).