8. TECHNICAL SPECIFICATIONS 

8.1. TECHNICAL 

CHARACTERISTICS 

Accuracy is referred to a temperature of 23°C±5°C with relative humidity <80%RH. 
 

Measurement of sound pressure level (SPL) 

Range [dB] 

Dynamic range 

Resolution 

Accuracy 

Frequency range 

Lo 

30 

 80 

50dB 0.1dB 

1.4dB 31.5Hz 

 8kHz 

Med 

50 

 100 

Hi 

80 

 130 

Auto 

30 

 130 

 

Integration 

Function Integration 

time 

FAST 125ms 

SLOW 1s 

 

Analogue output 

Type Output 

voltage 

Output 

impedance 

AC 

1Vrms with respect to selected FS 

100

 

DC 10mV/dB 

1k

 

 

Microphone 

Microphone with ½

” 

Electret condenser 

 

Reference standards 

Noise measurements: 

IEC61672-1 Class 2 

EMC: IEC/EN61326-1 

 

 

8.2. GENERAL 

CHARACTERISTICS 

Mechanical characteristics 

Size (L x W x H): 

260 x 65 x 50mm (10 x 3 x2in) 

Weight (battery included): 

280g (10ounces) 

Mechanical protection: 

IP40 

 

Power supply 

Battery type: 

1x9V alkaline battery type IEC 6F22 

Low battery indication: 

symbol 

“”

” 

on the display 

Battery duration: 

approx. 30 hours 

External power supply: 

100-240VAC, 50/60Hz, 0.3A / 9VDC, 500mA 

Auto Power OFF: 

after 15 minutes' idling 

 

Display 

Characteristics: 4 

LCD, Custom, backlit, bargraph 

Display update: 

2/s 

 

Recordings

 

selectable sampling interval between 1s and 59s 

 

max 32700 recordings can be saved in the internal memory 

 

Output connector

 mini 

USB

 

 

Содержание HTA102

Страница 1: ...Copyright HT ITALIA 2018 Version EN 1 00 17 04 2018 ENGLISH User manual...

Страница 2: ...TRUCTIONS 9 5 1 Real time measurement of sound pressure level 9 5 2 Recording of sound pressure level 9 5 3 Connection to analogue AC DC output 11 5 4 Calibration of the instrument 12 6 USING SOFTWARE...

Страница 3: ...ss 2 double insulation This instrument complies with the standards relevant to CE marking 1 1 PRELIMINARY INSTRUCTIONS We recommend following the normal safety rules devised to protect the instrument...

Страница 4: ...software SoundLink for recording operations On the instrument s front panel there are function keys to select the above mentioned functions see 4 3 The measured value appears on the wide LCD display...

Страница 5: ...cated in 8 4 1 In case of discrepancy please contact the Dealer In case the instrument should be returned please follow the instructions given in 9 3 2 INSTRUMENT POWER SUPPLY The instrument is suppli...

Страница 6: ...pod insertion 18 Battery compartment Fig 1 Description of the instrument 4 2 DESCRIPTION OF THE DISPLAY CAPTION 1 Low value indication 2 Active MIN function 3 Active FAST function 4 Full scale indicat...

Страница 7: ...isplay 4 3 5 Key LEV Press key LEV to select measuring range see Fig 2 part 6 The following ranges are available 30 80dB low 50 100dB medium 80 130dB high 30 130dB Autorange Message AUTO see Fig 2 par...

Страница 8: ...play Fig 3 System time setting 3 Press key SET to enter the programming section of Minutes The screen in Fig 3 Middle appears on the display 4 Press key LEV to set the Minutes and key SET to proceed a...

Страница 9: ...and exit the programming section OR 12 Press key HLD to reset In this case system date time is automatically set back to default configuration by the instrument This condition must be used in case it...

Страница 10: ...comfort use the provided tripod by inserting it into the relevant threaded hole on the instrument s back side see Fig 1 part 17 9 For functions HOLD and MAX MIN see 4 3 2 and 4 3 7 5 2 RECORDING OF SO...

Страница 11: ...vided external prower supply to the instrument s input see Fig 1 part 13 14 Looking at the indicated system time see Fig 2 part 16 press key REC to activate recording in the desired instant Message RE...

Страница 12: ...t using key ON OFF 5 3 CONNECTION TO ANALOGUE AC DC OUTPUT The instrument can be connected to external multimeters and or dataloggers for recording through the analogue AC DC output found on the instr...

Страница 13: ...y insert the mobile calibrator SC05 provided into the instrument s microphone until a click is heard as shown in Fig 11 4 Switch on the calibrator by selecting level 94dB at sinusoidal frequency 1kHz...

Страница 14: ...ure 3 Install USB driver by launching files CP210XCPInstaller_x86 exe 32 bit systems or CP210XCPInstaller_x64 exe 64 bit systems found on the CD ROM 4 Use ports COM3 or COM4 to connect the instrument...

Страница 15: ...the battery CAUTION Only expert technicians should perform this operation Before carrying out this operation make sure you have removed the probe from the input terminal 1 Switch off the instrument 2...

Страница 16: ...Electret condenser Reference standards Noise measurements IEC61672 1 Class 2 EMC IEC EN61326 1 8 2 GENERAL CHARACTERISTICS Mechanical characteristics Size L x W x H 260 x 65 x 50mm 10 x 3 x2in Weight...

Страница 17: ...000m 6562ft This instrument complies with Directive EMC 2014 30 EU This instrument satisfies the requirements of European Directive 2011 65 EU RoHS and 2012 19 EU WEEE 8 4 ACCESSORIES 8 4 1 Accessorie...

Страница 18: ...quence of improper packaging Repairs which may become necessary as a consequence of interventions performed by unauthorized personnel Modifications to the instrument performed without the manufacturer...

Отзывы: