HTA102
EN - 12
5.4.
CALIBRATION OF THE INSTRUMENT
Fig. 11: Connect the mobile calibrator to the instrument
1. Turn on the instrument by means of key
ON/OFF
2. Set the following configurations:
Weighting curve: A (dBA)
Integration type: FAST
Measuring range: 50 ÷ 100dB
3. Carefully insert the mobile calibrator
SC05
provided into the instrument
’
s microphone
until a click is heard, as shown in Fig. 11.
4. Switch on the calibrator by selecting level
94dB
at sinusoidal frequency 1kHz.
5. Insert the provided screwdriver into the trimmer (see Fig. 1
–
part 16) and carry out a
manual adjustment until the result
“
94.0
”
is shown on the display.
6. Extract the calibrator from the microphone and switch it off by moving the relevant
selector to the
OFF
position.