background image

Page  17

HUGO SACHS ELEKTRONIK - HARVARD 

A P P A R A T U S

 GmbH    D-79232 March-Hugstetten   Germany

HSE PLUGSYS

®

  TAM-A Type 705/1 and TAM-D Type 705/2

7.   Zero compensation ZERO

Zero compensation is required to correct the transducer zero without any load on. The output signal of e.g. a
Wheatstone bridge without any loading is never exactly zero. Possible causes are manufacturing tolerances in
the branch resistances, the effect of position, or slight bridge unbalance through overload or compensation of an
applied preload. Zero compensation consists of feeding a compensating current into the transducer bridge
circuit. In this way zero compensation is independent of the amplification selected on the TAM. TAM offers
manual coarse and fine adjustment as well as a very precise autozero function in order to compensate even
extreme symmetry errors in a transducer bridge.

7.1  Manual zero adjustment

The effect of the COARSE zero adjustment is larger than that of the FINE adjustment by a
factor of 5. In order to use the adjustment range of the FINE adjustment in both positive and
negative directions it is necessary first to move the FINE adjustment roughly to its centre
position. Turn the multi-turn adjustment fully anticlockwise (19 turns) and then about 8 turns clockwise. The same
applies to the autozero function, see description below (Section 7.2).

Coarse adjustment with the ZERO COARSE adjustment

Fine adjustment with the ZERO FINE adjustment

Precise zero with the autozero function

N o t e : On TAM-D it is in principle also possible to make the precise zero adjustment

using the built-in digital display and the FINE manual adjustment. It is therefore
not essential to produce precise zero adjustment using the autozero function. On
TAM-A however, accurate zeroing is not possible in this way because of the limited
resolution of the bargraph. Precise zero adjustment on TAM-A requires the use of
the autozero function.

7.2  Automatic zero adjustment.

Automatic zero compensation (autozero function) operates digitally with a resolution of 12 bit.
This provides an adjustment range of 4096 steps (±2048 steps). Zero adjustment operates
directly on the amplifier input and the zero shift per step therefore depends on the internal
resistance of the transducer and on the gain setting (see Section 4.2.5, Zero coupling). The
autozero setting is stored permanently and remains held until the next automatic zero adjustment even if the
amplifier case is switched off.

Automatic zero adjustment is possible only in operating mode MEASURE. Starting autozero
in the mode OFF or CAL results in an error, the status LED above the AUTO key flashes briefly 4 times.

R

Autozero can only be initiated after a waiting time of 2 seconds. This means that the
AUTO push button must be held down for at least 2 seconds. This prevents unintentional
initiation of a zero adjustment. Throughout the adjustment the status LED shows a
steady light. After successful zero adjustment the status LED goes off immediately. In case the autozero
adjustment range is not sufficient, the status LED flashes 15 times at the end of the adjustment. In
addition the automatic system is normalised to the centre of the adjustment range. In order to proceed
with zeroing the zero position has to be corrected manually. Automatic adjustment should then be possi-
ble (see also autozero coupling Section 4.2.5).

R

In order to move autozero to its mid position the transducer has to be disconnected from the TAM and
autozero is started. Since zeroing is now impossible the autozero function makes an unsuccessful attempt
at zeroing and then normalises itself to its mid position.

R

Содержание PLUGSYS TAM-A

Страница 1: ...on 1 4 printet June 2013 Hellstab TB Zi OPERATING INSTRUCTIONS TAM D HUGO SACHS ELEKTRONIK D 79232 MARCH GERMANY TRANSDUCER AMPLIFIER MODULE CAL MEASURE OFF 1 300 100 FILTER Hz GAIN 20 100 OUTPUT 10V TYPE 705 2 AUTO FINE COARSE ZERO REF TRANSDUCER H SE ADJ 1 3 5 2 4 6 300 1 3 5 2 4 6 FILTER Hz TAM A HUGO SACHS ELEKTRONIK D 79232 MARCH GERMANY TRANSDUCER AMPLIFIER MODULE CAL MEASURE OFF 1 100 GAIN ...

Страница 2: ... jumper J3 9 4 2 7 BNC socket Puls or mean output jumper J4 9 4 2 8 Internal signal outputs PULSE and MEAN jumper J5 9 4 2 9 Setting of the decimal point TAM D jumper J1 10 4 2 10 Mean value filter digital display TAM D jumper J6 10 4 3 Signal input 10 5 Controls TAM A 11 5 1 Controls TAM D 12 6 Start up 13 6 1 Transducer calibration 13 6 1 1 Basic setting for transducer calibration 13 6 1 2 Trans...

Страница 3: ...d for use in general laboratories light industrial and office environments Operation in hazardous areas and or in a flammable atmosphere is not permitted 1 2 Trademark Warning All the information in these Instructions has been drawn up after careful examination but does not represent a warranty of product properties Alterations in line with technical progress are re served These Operating Instruct...

Страница 4: ...TAM B are equipped with automatic zeroing in addition to manual zero adjustment In order to have an easy calibration of a connected data acqusistion system a predefinded value eg 10mN 100mmHg or 5mm can be simulated on the output The output filter 30 100 300 Hz is used for smoothing the signal It acts on both signal outputs PULSE internal and BNC socket front panel The internal signal output MEAN ...

Страница 5: ...h the inner pins of the input or output sockets For this reason you should always first provide potential equilibration between yourself the PLUGSYS and the transducer or recording system 4 2 Internal settings jumpers In order to adjust the TAM to the individual application the following settings have to be made with jumpers on the module circuit board See diagrams next page R R R Jumper J1 TAM D ...

Страница 6: ...2 DP 3 DP 4 J1 SLOW FAST J6 POS 1 POS 2 0 1Hz 0 3Hz Filter MEAN PULSATILE MEAN OUTPUT front panel A low B mean C high RANGE of AUTO ZERO J5 In the circle asside you see 3 coulums of pins On all left pins the pulsatile signal is available The center pins are connected to the bus lines AV1 AV16 On all right pins you have the filtered mean signal avaialable To connect e g the pulsatile pressure signa...

Страница 7: ... the measurement signal is amplified by these active transducers up to the Volt range The input amplifier of the TAM is switched as straight amplifier Jumper J7 on POS 1 case shield Transducer Amp Output Power Power Sig GND Typical arrangement of a transducer with preamplifier In order to avoid damage to the electronics of the transducer check first that a supply voltage of 5 Volt 50 mA can be use...

Страница 8: ...r does only allow gain reduction the gain jumper should allways be set to the next higher possible value In the example above the gain should be set to 200 This gives an overall amplification of 200 preamplifier times 10 output amplifier equal to 2000x The exact adjustment of the gain to 1980 is made through the transducer calibration on the front trimmer GAIN 20 100 10 5 05 x 10 6 x 5 x 200 1980 ...

Страница 9: ...nternal signal outputs PULSE and MEAN jumper J5 The internal signal outputs PULSE and MEAN of the TAM are each assigned by a jumper on the link connector J5 to an analogue bus line AV1 AV16 Be sure to select an AV bus line which is not in use and is therefore free Duplicate connection causes short circuiting of the two signal outputs Details of the AV lines already being used in the system can be ...

Страница 10: ...value then corresponds exactly of the measuring range which you require For example for the range 10 00 mN the jumper should be set to position DP 3 In case no decimal point is required remove the jumper and set it to one pin For measuring blood pressures up to 200mmHg you have to remove this jumper so that no decimal point is displayed 4 3 Signal input The signal input is a 6 pin Binder or Amphen...

Страница 11: ... 300 100 GAIN 20 100 OUTPUT 10V TYPE 705 1 AUTO FINE COARSE ZERO REF TRANSDUCER 25 0 25 50 75 H SE 1 3 5 2 4 6 FILTER Hz LED bargraph Marker field for reference e g 10 mN Marker field for calibrated transducer e g Type F30 No 1 Signal input Output PULSE or MEAN Filter settings Operating mode CAL setting Amplification setting Autozero status LED Autozero key Manual zero fine Manual zero coarse 100 ...

Страница 12: ...ODULE CAL MEASURE OFF 1 300 100 FILTER Hz GAIN 20 100 OUTPUT 10V TYPE 705 2 AUTO FINE COARSE ZERO REF TRANSDUCER ADJ 1 3 5 2 4 6 Marker field for calibrated transducer e g Type F30 No 1 Signal input Output PULSE or MEAN Filter setting Operating mode CAL setting Amplification setting Autozero status LED Autozero key Manual zero fine Manual zero coarse LED digital display Display range adjustment Ma...

Страница 13: ...100 mmHg 6 1 1 Basic settings for transducer calibration R Set output filter to 300 Hz R Set operating mode to MEASURE R Set GAIN 20 100 to 100 fully clockwise clicking noises on the multi turn trimmer R Set fine manual zero adjustment to mid position 19 turns anticlockwise and then 9 turns clockwise Initialize the autozero function to the mid of the calibration range To do so start the autozero f...

Страница 14: ...e measuring range Tip R The final step should be to check the zero again The calibration reference load on the trans ducer is removed The signal output of the TAM must then again be exactly zero If this is not the case it is probable that application of the calibration reference load has shifted the trans ducer zero A possible cause is a change in the position of the transducer pulling on the conn...

Страница 15: ... ZERO COARSE first and then ZERO FINE Adjust Zero so that your voltmeter shows al most zero Detailed description R Start the AUTOZERO function by pressing the AUTO button in MEASURE MODE until the LED ZERO goes on for precise zeroing The LED ZERO must automatically go off after a few seconds and not be flashing at the end of the adjustment no error zero adjustment OK R Using a pressure calibrator ...

Страница 16: ...o CAL position Turn on the CAL trimmer so that your display shows the choosen simulation value e g 100 mmHg If you switch now to CAL your TAM D always simulates 100mmHg on the display and on the output R R 6 1 4 Transducer calibration on a TAM D with using the digital display procedure for users who don t have a digital voltmetern nor a DAQ system R R With no load on the transducer carry out an ap...

Страница 17: ...utozero function On TAM A however accurate zeroing is not possible in this way because of the limited resolution of the bargraph Precise zero adjustment on TAM A requires the use of the autozero function 7 2 Automatic zero adjustment Automatic zero compensation autozero function operates digitally with a resolution of 12 bit This provides an adjustment range of 4096 steps 2048 steps Zero adjustmen...

Страница 18: ...00 Hz 9 Maintenance notes Any splashes of salt solution should be removed immediately with a cloth in order to prevent corrosion damage to the metal parts the connectors and the electronics For cleaning the front panel connectors and cables never use scouring powder or cleaning agents which attack plastics or aluminium Any dust should be removed with a lint free cloth or with a fine dust brush Hea...

Страница 19: ...y internal jumper as x10 x20 x50 x100 x200 x500 x1000 x2000 x5000 and x10000 The basic gain setting can be varied with the GAIN control within the range 20 100 Gain 1 Autozero by pushbutton 2 Manual zero through multi turn coarse and fine trimmers Zeroing Analogue LED bargraph 19 LEDs 12 null 6 for visual indication of the measurement signal Indication TAM A 3 digit LED display 10 Volt 100 or vari...

Страница 20: ...F Filter 1 100 300 Hz 17 Filter setting mean output 9 G General description 4 I Input configuration of the amplifier 7 Installation procedure 4 Installing the module in a housing 4 Internal settings 5 Internal signal outputs PULSE MEAN 9 Introduction 3 J Jumper location of 6 M Maintenance notes 17 Manual zero adjustment 16 Manufacturer s details 3 Mean value filter digital display TAM D 10 Mean va...

Страница 21: ...elp to improve our products and the system documentation and make them more user friendly Please tell us where you have found mistakes where the arrangement was not clear and what you did not understand and where you would like to see improvements Many thanks for your kind assistance Yours HUGO SACHS ELEKTRONIK HARVARD APPARATUS GmbH Your name Organisation Street Town Phone Fax e mail Please send ...

Страница 22: ...ssing the AUTO ZERO button on the TAM module for 3 seconds The red LED light above the button lights up during automatic zero adjustment If the module can not adjust zero the red LED light starts blinking first is to check if switch CAL OFF MEASURE is really in MEASURE position If the module can not adjust Zero you have to adjust Zero manually with the COARSE trimmer below to the AUTO ZERO button ...

Страница 23: ...t panel with the little red screwdriver which was delivered with the system adjust GAIN so that you get 5V 50 bargraph deflection for 100mmHg The output delivers the voltage of 5V remove pressure from your transducer you should get 0V on the output again give 100mmHg to the transducer you should receive 5V on the output if this is the case all is set well If not readjust In order to avoid performi...

Страница 24: ...0 Hz Move the switch CAL OFF MEASURE in MEASURE position First step is the Zero adjustment Remove load weight from the transducer Adjust Zero by pressing the AUTO ZERO button on the TAM module for 3 seconds The red LED light above the button lights up during automatic zero adjustment If the module can not adjust zero the red LED light starts blinking first is to check if switch CAL OFF MEASURE is ...

Страница 25: ...and on the output socket of the module Take the little red screwdriver and turn on the CAL trimmer not on the GAIN so that you get 1V 10 bargraph deflection which is equal to 1g Now all settings are done and you can switch back the module to MEASURE position and start your measurements If you e g need more accuracy but less range you can increase the GAIN e g by a factor of 2 1g 2V Range will be h...

Отзывы: