EMC Notice
Electromagnetic emissions
The HC271/HC271p monitors are intended for use in the electromagnetic environment specified below. The
customers or the users of the HC271/HC271p monitors should assure that they are used in such an
environment.
Table 1-3
Electromagnetic emissions
Emissions test
Compliance
Electromagnetic environment - Guidance
RF emissions
CISPR 11
Group 1
The HC271/HC271p uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B
The HC271/HC271p is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000–3–2
Class D
Voltage fluctuations/ flicker emissions
IEC 61000–3–3
Complies
Electromagnetic immunity
Table 1-4
Electromagnetic immunity
Immunity test
IEC 60601–1–2:2014 Test
levels
Compliance level
Electromagnetic environment
— Guidance
Electrostatic discharge (ESD)
IEC 61000–4–2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Electrical fast transient/burst
IEC 61000–4–4
±2 kV for power supply lines
±2 kV for power supply lines
The main power quality should
be that of a typical commercial
or hospital environment.
Surge
IEC 61000–4–5
±1 kV line(s) to line(s)
±2 kV line(s) to earth
±1 kV line(s) to line(s)
±2 kV line(s) to earth
The main power quality should
be that of a typical commercial
or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000–4–11
< 5% U
T
1 (> 95% dip in U
T
) for
0.5 cycle
40% U
T
(60% dip in U
T
) for 5
cycles
70% U
T
(30% dip in U
T
) for 25
cycles
< 5% U
T
1 (> 95% dip in U
T
) for
250 cycles
< 5% U
T
1 (> 95% dip in U
T
) for
0.5 cycle
40% U
T
(60% dip in U
T
) for 5
cycles
70% U
T
(30% dip in U
T
) for 25
cycles
< 5% U
T
1 (> 95% dip in U
T
) for
250 cycles
The main power quality should
be that of a typical commercial
or hospital environment. If the
user of the HC271/HC271p
requires continued operation
during interruptions in the main
power, it is recommended that
the HC271/HC271p be powered
from an uninterruptible power
supply or a battery.