Power Meter Remote Operation
Getting the Best Speed Performance
1-66
HP E4418B/E4419B Programming Guide
Getting the Best Speed Performance
This section discusses the factors that influence the speed of operation
(number of readings/sec) of an HP E4418B/E4419B power meter.
The following factors are those which have the greatest effect upon
measurement speed (in no particular order):
•
The selected speed i.e. 20, 40 or 200 readings/sec.
•
The trigger mode (for example, free run, trigger with delay etc.).
•
The output format i.e. ASCii or REAL.
•
The units used for the measurement.
•
The command used to take a measurement.
In addition, in 200 reading/sec mode there are other influences which are
described in “200 Readings/Sec”, on page 1-68.
The following paragraphs give a brief description of the above factors and
how they are controlled from SCPI.
Speed
There are three possible speed settings 20, 40 and 200 readings/sec. These
are set using the
SENSe:SPEed
command and can be applied to each
channel independently (HP E4419B only). The speed setting controls the
cycle time of the measurement i.e., 50ms, 25ms and 5ms respectively.
In 20 and 40 readings/sec mode, full instrument functionality is available;
200 readings/sec is available only for E-series sensors and averaging,
offsets, limits, and ratio/difference math functions are disabled.
Refer to “Specifications” in chapter 5 of the User’s Guide to see the
influence of these speed settings on the accuracy and noise performance of
the power meter.
Trigger Mode
The power meter has a very flexible triggering system. For simplicity, it
can be described as having three modes:
Содержание E4418B
Страница 22: ...prog book progLOF doc 12 Mon Feb 15 09 24 37 1999 Contents 12 HP E4418B E4419B Programming Guide ...
Страница 24: ...prog book progLOT doc 14 Mon Feb 15 09 24 37 1999 Contents 14 HP E4418B E4419B Programming Guide ...
Страница 25: ...prog book chapter1 fm 1 Mon Feb 15 09 24 37 1999 1 Power Meter Remote Operation ...
Страница 138: ...prog book ch2meas fm 1 Mon Feb 15 09 24 37 1999 2 MEASurement Instructions ...
Страница 200: ...prog book ch3calc fm 1 Mon Feb 15 09 24 37 1999 3 CALCulate Subsystem ...
Страница 228: ...prog book ch4cal fm 1 Mon Feb 15 09 24 37 1999 4 CALibration Subsystem ...
Страница 242: ...prog book ch5disp fm 1 Mon Feb 15 09 24 37 1999 5 DISPlay Subsystem ...
Страница 260: ...prog book ch6form fm 1 Mon Feb 15 09 24 37 1999 6 FORMat Subsystem ...
Страница 264: ...prog book ch7mem fm 1 Mon Feb 15 09 24 37 1999 7 MEMory Subsystem ...
Страница 294: ...prog book ch8out fm 1 Mon Feb 15 09 24 37 1999 8 OUTput Subsystem ...
Страница 302: ...prog book ch9sens fm 1 Mon Feb 15 09 24 37 1999 9 SENSe Subsystem ...
Страница 360: ...prog book ch10stat fm 1 Mon Feb 15 09 24 37 1999 10 STATus Subsystem ...
Страница 384: ...prog book ch11syst fm 1 Mon Feb 15 09 24 37 1999 11 SYSTem Subsystem ...
Страница 420: ...prog book ch12trig fm 1 Mon Feb 15 09 24 37 1999 12 TRIGger Subsystem ...
Страница 434: ...prog book ch13unit fm 1 Mon Feb 15 09 24 37 1999 13 UNIT Subsystem ...
Страница 442: ...prog book ch14serv fm 1 Mon Feb 15 09 24 37 1999 14 SERVice Subsystem ...
Страница 452: ...prog book ch15ieee fm 1 Mon Feb 15 09 24 37 1999 15 IEEE488 2 Command Reference ...
Страница 481: ...prog book progIX doc 6 Mon Feb 15 09 24 37 1999 Index 6 HP E4418B E4419B Programming Guide ...