Check the activity indicators for status lines locked in a high or low
state.
Verify that the STAT, DATA, and ADDR format labels have not been
modified from their default values.
Except in certain cases (see page 1-7), these labels must remain as they are
configured by the configuration file. Do not change the names of these labels
or the bit assignments within the labels. Some preprocessors also require
other data labels. See Chapter 2 for more information.
Verify that all microprocessor caches and memory managers have been
disabled.
In most cases, if the microprocessor caches and memory managers remain
enabled you should still get inverse assembly. It may be incorrect because a
portion of the execution trace was not visible to the logic analyzer.
Verify that storage qualification has not excluded storage of all the
needed opcodes and operands.
Inverse assembler will not load or run
You need to ensure that you have the correct system software loaded on your
analyzer.
Ensure that the inverse assembler is on the same disk as the
configuration files you are loading.
Configuration files for the state analyzer contain a pointer to the name of the
corresponding inverse assembler. If you delete the inverse assembler or
rename it, the configuration process will fail to load the disassembler.
See Chapter 1 for details.
If You Have a Problem
Inverse assembler will not load or run
A–8
Содержание E2470A
Страница 4: ...HP E2470A Preprocessor Interface Introduction The HP E2470A Preprocessor Interface At a Glance iii ...
Страница 8: ...1 Setting Up the Preprocessor Interface ...
Страница 22: ...2 Analyzing the Motorola MC68HC16Y1 ...
Страница 34: ...3 Preprocessor Interface Hardware Reference ...
Страница 38: ...Block Diagram Preprocessor Interface Hardware Reference Theory of Operation and Clocking 3 5 ...
Страница 56: ...A If You Have a Problem ...