Chapter 5
Adjustments
Introduction
This chapter contains procedures to adjust the HP E1411 multimeter for
peak performance. For best performance, the instrument should be adjusted
after repair. All adjustments are performed electrically, so manual
adjustment of the multimeter is not necessary.
WARNING
Do not perform any of the following adjustments unless you are
a qualified, service-trained person, and have read the
WARNINGS and CAUTIONS in
Chapter 1 - General Information.
NOTE
ALL adjustment procedures MUST be performed in the order shown in this
manual (DC Voltage, then AC Voltage, and then Resistance).
Adjustment
Conditions /
Procedures
For valid adjustments, the HP E1411 multimeter and test equipment used
must have at least a 60 minute warm-up, and the line voltage must be
115/230 Vac
±
10%. For best accuracy, the temperature of the area where
adjustments are made should be between 18
o
C and 28
o
C and stable to
within
±
1
o
C. See Table 1-1, Recommended Test Equipment, for test
equipment requirements.
The adjustment procedures assume the person performing the adjustments
understands how to operate the mainframe, multimeter and specified test
equipment. The adjustment procedures do not specify test equipment
settings, except in general terms. It is assumed a qualified, service-trained
person will select and connect the cables and jumpers required for the
adjustments.
Chapter 5
Adjustments 37
Содержание E1411A
Страница 4: ...Notes ...
Страница 8: ...Notes 8 HP E1411A B Service Manual ...
Страница 10: ...10 HP E1411A E1411B Service Manual ...
Страница 36: ...Notes 36 Verification Tests Chapter 4 ...
Страница 48: ...Notes 48 Adjustments Chapter 5 ...
Страница 51: ...Figure 6 1 HP E1411A Replaceable Parts Chapter 6 Replaceable Parts 51 ...
Страница 53: ...Figure 6 2 HP E1411B Replaceable Parts Chapter 6 Replaceable Parts 53 ...
Страница 56: ...56 Manual Changes Chapter 7 ...