230 END
Test 4-4: Resistance Test (4-Wire Ohms)
This test verifies the 4-wire resistance accuracy of the 2k
Ω
, 100k
Ω
, and
1M
Ω
ranges.
NOTE
The DC Voltage performance test must be performed prior to the Resistance
Test to check the A/D accuracy on all ranges. If the DC Voltage test has not
been performed, resistance must be checked on all ranges at 0 and at 50%
of full scale.
Equipment Setup
1.
Connect the equipment as shown in Figure 4-4
2.
Set the HP E1411 multimeter as follows:
•
Reset Multimeter ..................................................*RST
•
Auto Zero ..................................................................ON
•
Power Line Cycles (PLC) .............................................1
Figure 4-4. Resistance Test Setup
Chapter 4
Verification Tests 29