WARNING
POSSIBLE ELECTRICAL SHOCK. Calibration and performance tests are to be performed by
qualified personnel only. Do not attempt calibration or test procedures unless qualified to do so.
CAUTION
To avoid damage to the multimeter for inputs above 250 Vdc or Vac, disconnect the test leads
before changing functions. Do not exceed the maximum input limits shown in the following table.
Maximum Overvoltage Limitations (AC and DC Voltage Functions)
1000V
MAX
indicates the maximum voltage between input terminals and earth is
±
1000 V (dc or ac rms).
Do not use the multimeter on any ACV circuit where the maximum impulse overvoltage may
be more than 4000Vpk or any DCV circuit where the maximum impulse overvoltage may be
more than 2500Vpk between the COM and VOLT terminals. Excessive impulse overvoltage
can damage the multimeter voltage functions. Do not measure branch circuits (CAT II) over
600V to earth nor service panel circuits (CAT III) over 300V to earth.
Safety Summary
1 - 5