Adjustments
Calibration Table
CAUTION
Dangerous voltages are present during the calibration procedure. Calibration should only be
performed by qualified service technicians. Use a non-conductive tool.
Step
Function
Range
Input Signal
Adjustment
(limits)
Tolerance (counts)
972A
973A
1
mV
400 mV
Short
—
±
1
±
1
2
380.0 mV
1 (
±
1)
±
8
±
4
3
40 mV
38.00 mV
2 (
±
1)
±
18
±
16
4
V
400 V
380.0 V
3 (
±
1)
±
8
±
4
5
4 V
3.800 V
4 (
±
1)
±
8
±
4
6
40 V
38.00 V
5 (
±
1)
±
8
±
4
973A
972A
6 - 1