AC Voltage (Average responding, calibrated to display rms)
Range
Resolution
Accuracy
Input Impedance (nominal)
40 Hz to 500 Hz
500 Hz to 1 kHz
400 mV
0.1 mV
±
(1% + 2)
±
(1.5% + 4)
10 M
Ω
< 50 pF
4 V
1 mV
11 M
Ω
< 50 pF
40 V
10 mV
10 M
Ω
< 50 pF
400 V
100 mV
1000 V
1 V
Not Specified
Common Mode Rejection Ratio (CMR)1 k
Ω
imbalance: > 60 dB @ DC to 60 Hz
Response time: 2 seconds maximum
DC Current
Range
Resolution
Accuracy
Input
Resistance
Maximum Input
400
µ
A
100 nA
±
(0.5% + 2)
< 550
Ω
±
0.5 A (fused)
4000
µ
A
1
µ
A
±
(1% + 2)
40 mA
10
µ
A
< 8
Ω
400 mA
100
µ
A
±
(1.2% + 2)
10 A
10 mA
< 0.05
Ω
±
15 A (fused)