![HP 971A Скачать руководство пользователя страница 21](http://html.mh-extra.com/html/hp/971a/971a_user-manual_162640021.webp)
Diode
Measurement current: +0.5 mA nominal @ 0.6 V Open circuit voltage: < 3.2 Vpeak
Displayed Voltage: 0 V to 2.000 V
Input protection: 660 Vrms (sinewave)
Accuracy:
±
(1% + 2)
Resolution: 1 mV
Temperature (5 k
Ω
@ 25
°
C Thermistor probe)
°
C
°
F
Measurement
Range
-80
°
to 150
°
-112
°
to 302
°
Resolution
0.1
°
0.2
°
Accuracy
1
±
0.5
°
±
1.0
°
1
Accuracy does not include 5 k
Ω
Thermistor error
Frequency (AC Coupled)
Frequency Range
Resolution
Accuracy
Maximum Input
5 Hz to 9999 Hz
1 Hz
±
(0.02% + 1)
660 Vrms
9.0 kHz to 99.9 kHz
10 Hz
100 Vrms
Input
Attenuator
Input Voltage
Input Impedance
5 Hz to 10 kHz
10 kHz to 100 kHz
4
100 mV to 4 V rms
800 mV to 4 V rms
2.3 M
Ω
< 50 pF
40
4 V to 40 V rms
4 V to 40 V rms
1.3 M
Ω
< 50 pF
400
40 V to 400 Vrms
40 V to 100 V rms
1.2 M
Ω
< 50 pF
Specifications