24
Accessories
Calibration kits
Vector accuracy enhancement procedures require that
the systematic errors of the measurement system be
characterized by measuring known devices (standards)
on the system over the frequency range of interest.
Return loss specifications or typical values are provid-
ed where available for the terminations and adapters.
HP 85031B 7-mm calibration kit
Contains precision 7-mm standards used to calibrate
the HP 8753ES for measurement of devices with preci-
sion 7-mm connectors.
Includes
HP part number
7 mm short/open circuit
85031-60001
7 mm 50-ohm terminations
(two each)
00909-60008
Specifications for terminations
DC to 5 GHz
RL
≥
52 dB
5 to 6 GHz
RL
≥
46 dB
HP 85032B 50-ohm type-N calibration kit
Contains precision 50-ohm type-N standards used to
calibrate the HP 8753ET/ES and 50-ohm test sets for
measurement of devices with 50-ohm type-N connec-
tors. Precision phase-matched 7-mm to type-N
adapters are included for accurate measurements of
non-insertable devices. Option 001 deletes the 7-mm
to type-N adapters, which are not needed with the
HP 8753ET.
Includes
HP part number
N-male 50-ohm termination
00909-60009
N-female 50-ohm termination
00909-60010
N-male short circuit
85032-60008
N-female short circuit
85032-60009
N-female open circuit
85032-60012
N-male open circuit
85032-60007
7-mm to N-male adapter
(two each)
85054-60009
7-mm to N-female adapter
(two each)
85054-60001
Specifications for terminations:
DC to 3 GHz
RL
≥
49 dB
2 to 3 GHz
RL
≥
46 dB
3 to 6 GHz
RL
≥
40 dB
Typical adapter characteristics:
DC to 6 GHz
RL
≥
30 dB
HP 85033D 3.5-mm calibration kit
Contains a set of precision 3.5-mm standards to cali-
brate the HP 8753ES and 50-ohm test sets for the
measurement of devices with precision 3.5-mm and
SMA connectors. Precision phase-matched 7-mm to
3.5-mm adapters are included for accurate measure-
ments of non-insertable devices. Option 001 deletes
the 7-mm to type-N adapters.
Includes
HP part number
3.5-mm-male 50-ohm termination
85033-60009
3.5-mm-female 50-ohm termination 85033-60010
3.5-mm-female short
85033-60014
3.5-mm-male short
85033-60013
3.5-mm-female open
85033-60012
3.5-mm-male open
85033-60011
7-mm to 3.5-mm female
adapter (two)
1250-1747
7-mm to 3.5-mm male
adapter (two)
1250-1746
Specifications for terminations:
DC to 1.3 GHz
RL
≥
46 dB
1.3 to 3 GHz
RL
≥
44 dB
3 to 6 GHz
RL
≥
38 dB
Typical adapter characteristics:
DC to 6 GHz
RL
≥
34 dB
HP 85036B 75-ohm type-N calibration kit
Contains a set of precision 75-ohm type-N standards to
calibrate the HP 8753ES with a 75-ohm test set for
measurement of devices with 75-ohm type-N connec-
tors. Precision phased-matched adapters are included
for accurate measurements of non-insertable devices.
Includes
HP part number
N-male 75-ohm termination
00909-60019
N-female 75-ohm termination
00909-60020
N-female 75-ohm short
85036-60011
N-male 75-ohm short
85036-60012
N-female open
85032-20001
N-male open
85032-60007
N-male to N-male 75-ohm adapter
85036-60013
N-female to N-female 75-ohm adapter 85036-60014
N-male to N-female 75-ohm adapter
85036-60015
Specifications for terminations:
DC to 2 GHz
RL
≥
46 dB
2 to 3 GHz
RL
≥
40 dB