B-4
HP 85325A Operating and Service Manual
Connecting the Waveguide Properly
Tighten Flange
Connections
8. In an X pattern (for equal compression), tighten all four screws (see
figure below). Do not over-tighten.
9. Visually inspect the connection.
Figure B-4
Inspect Flange Connections
NOTE
The best connection has symmetrical pressure applied as you gradually
tighten the screws.
Inspect Flange
Connections
Visually inspect a flange connection as follows:
10. Place an electric light or white paper behind the connection.
11. Check the flange matings for any gap. A good connection has no gaps
between the connected waveguide flanges, and the waveguide walls are
flush (there is no step or offset).
12. Make sure that all four screws are equally torqued.
NOTE
Remember the most common cause of measurement error is a poor
connection.