
HP 8517B S-Parameter Test Set Manual
3-12
5. View the S
21
measurement. If the thru is connected, the transmission
coefficient should be 1
∠
0
°
.
6. Press PARAMETER
[S
12
]
, then press the following
DISPLAY:
menu keys:
{SELECT DEFAULTS}
{DEFAULT to MEMORY: 2}
7. Press PARAMETER
[S
12
]
to view the S
12
measurement. If the thru is
connected, the transmission coefficient should be 1
∠
0
°
.
Normalization is turned OFF for S
11
and S
22
measurements. Then the
appropriate memory is selected and normalization is turned on to view S
21
and S
12
.
Since these are accurate, repeatable attenuators, this sequence may be
effective in your application. If other parts of the test must be changed,
especially components connected to the test ports, this procedure may
require additional steps to adequately compensate for the changes.
Содержание 8517B
Страница 10: ...x HP 8517B S Parameter Test Set Manual ...
Страница 16: ...Contents 6 HP 8517B S Parameter Test Set Manual ...
Страница 30: ...1 14 HP 8517B S Parameter Test Set Manual Getting Started Operating and Safety Precautions ...
Страница 38: ...2 8 HP 8517B S Parameter Test Set Manual Installation Packaging the Test Set ...
Страница 57: ...HP 8517B S Parameter Test Set Manual 1 ...
Страница 60: ...HP 8517B S Parameter Test Set Manual 4 4 Specifications ...
Страница 84: ...5 24 HP 8517B S Parameter Test Set Manual Troubleshooting the Test Set Troubleshooting Procedures ...
Страница 109: ...HP 8517B S Parameter Test Set Manual 7 11 HP 8517B Replaceable Parts Figure 7 1 Power Cable and Plug Part Numbers ...
Страница 121: ...HP 8517B S Parameter Test Set Manual 7 23 HP 8517B Replaceable Parts Figure 7 4 Detailed Views of Hardware Stack Ups ...