
HP 8517B S-Parameter Test Set Manual
3-7
Figure 3-4
RF and IF Switching with Four Test Sets
Table 3-5
Coaxial Switch Settings for Figure 3-4, Above
New ADDRESS
of Test Set
1,
2, 3
1. Not all system connections are shown.
2. In dual source configurations, the second source can be multiplexed in a similar manner.
3. If only one dual source test set is used, the second source can be directly connected to the
appropriate test set.
Test Set Selected
1, 2, 3
Coaxial Port Selected
with Switch #1
Coaxial Port Selected
with Switch #2
20
1
Port 1
Port 1
21
2
Port 1
Port 2
22
3
Port 2
Port 1
23
4
Port 2
Port 2
Содержание 8517B
Страница 10: ...x HP 8517B S Parameter Test Set Manual ...
Страница 16: ...Contents 6 HP 8517B S Parameter Test Set Manual ...
Страница 30: ...1 14 HP 8517B S Parameter Test Set Manual Getting Started Operating and Safety Precautions ...
Страница 38: ...2 8 HP 8517B S Parameter Test Set Manual Installation Packaging the Test Set ...
Страница 57: ...HP 8517B S Parameter Test Set Manual 1 ...
Страница 60: ...HP 8517B S Parameter Test Set Manual 4 4 Specifications ...
Страница 84: ...5 24 HP 8517B S Parameter Test Set Manual Troubleshooting the Test Set Troubleshooting Procedures ...
Страница 109: ...HP 8517B S Parameter Test Set Manual 7 11 HP 8517B Replaceable Parts Figure 7 1 Power Cable and Plug Part Numbers ...
Страница 121: ...HP 8517B S Parameter Test Set Manual 7 23 HP 8517B Replaceable Parts Figure 7 4 Detailed Views of Hardware Stack Ups ...