cycle
is
complete.
This
information
is
imp ortant
for
sync
hronizing
the
pulse
mo
dulation
to
the
external
trigger.
Shen
a
TTL
pulse
train
is
applied
to
the
HP
8510
rear
panel
EXTERNAL
TRIGGER
INPUT
the
trigger
should
stay
low
for
at
leas
100
nanoseconds.
Excess
triggers
that
o ccur
while
stop
sw
eep
is
busy
(low)
are
ignored.
The
next
external
trigger
falling
edge
after
stop
sw
eep
is
ready
will
initiate
the
next
measuremen
t
cycle
and
retrace
is
automatic.
Figure
3-21.
External
Trigger
and
Stop
Sw
eep
Timing
Diagram
When
in
ternal
triggering
is
selected,
the
stop
sw
eep
output
do es
not
op erate
(alw
a
ys
in
the
ready
state).
Timing
diagrams
for
the
pulse
output
signal
and
dev
elopmen
t
of
the
in
ternal
measuremen
t
trigger
are
sho
wn
in
Figure
3-11
and
Figure
3-12 .
When
external
triggering
is
selected,
the
falling
edge
of
the
external
trigger
input
denes
time
equals
zero
seconds
for
eac
h
measuremen
t
cycle.
Stop
sw
eep
falls
immediately
and
stays
low
un
til
the
HP
8510
completes
the
measuremen
t
cycle
and
is
ready
to
accept
the
next
trigger.
The
time
p erio
d
that
stop
sw
eep
remains
busy
dep ends
up on
the
next
measuremen
t
function
to
b e
p erformed.
Pulse
output
is
turned
O
for
external
triggering.
As
for
in
ternal
triggering,
the
pulse
prole
measuremen
t
resolution
p erio
d
is
set
b
y
the
larger
of
pulse
width
and
pulse
prole
domain
stop
time.
Using
External
Triggering
and
Pulse
Modulation
In
applications
where
it
is
necessary
to
maintain
close
con
trol
o
v
er
the
PRP
and
dut
y
cycle
of
the
pulsed-RF
stim
ulus,
y
ou
can
use
external
equipmen
t
to
pro
vide
the
TTL
pulse
mo
dulation
and
external
trigger
signals.
When
using
an
external
pulse
mo
dulation
signal,
it
is
necessary
to
sync
hronize
the
net
w
ork
analyzer
with
the
pulse
mo
dulation
signal
so
that
the
measuremen
t
is
alw
a
ys
made
at
the
same
time
with
resp ect
to
the
stimulus.
This
sync
hronization
is
accomplished
using
the
HP
8510
external
trigger
input.
Connect
the
Pulse
Generator
A
simple
metho d
to
con
trol
the
PRP
and
dut
y
cycle
of
the
pulsed-RF
stimulus
is
to
use
an
external
pulse
generator
to
pro
vide
the
pulse
mo
dulation
input
to
the
RF
source.
Figure
3-22
sho
ws
a
simple
setup
that
pro
vides
the
same
pulsed-RF
stimulus
to
the
DUT
regardless
of
the
3-26
Operation
All manuals and user guides at all-guides.com
Содержание 85108A
Страница 4: ...All manuals and user guides at all guides com ...
Страница 18: ...All manuals and user guides at all guides com ...
Страница 84: ...All manuals and user guides at all guides com ...
Страница 85: ...All manuals and user guides at all guides com ...
Страница 86: ...All manuals and user guides at all guides com ...
Страница 92: ...All manuals and user guides at all guides com ...
Страница 112: ...6 20 System Service and Troubleshooting All manuals and user guides at all guides com ...
Страница 113: ...Figure 6 7 HP 85108A Block Diagram All manuals and user guides at all guides com ...
Страница 114: ...All manuals and user guides at all guides com ...
Страница 116: ...All manuals and user guides at all guides com ...
Страница 118: ...Figure 7 1 HP 85108 Replaceable Parts 7 2 Replaceable Parts All manuals and user guides at all guides com ...
Страница 121: ...Figure 7 2 HP 85108 Replaceable Parts Replaceable Parts 7 5 All manuals and user guides at all guides com ...
Страница 126: ...All manuals and user guides at all guides com ...
Страница 132: ...All manuals and user guides at all guides com ...
Страница 136: ...All manuals and user guides at all guides com ...