DRAFT
29
Operation and Service Manual
Performance Tests Procedures
Figure 7
Distortion and Reverse Isolation Test Setup
Procedure:
1
To check distortion, connect the equipment as shown in Figure 7. Set the sweep oscil-
lator for a fixed, 400 MHz CW signal.
For the HP 8447F option H64 low frequency preamp, set the frequency to 10 MHz.
2
Set the spectrum analyzer input attenuator as shown below. Adjust the controls to dis-
play 400 through 1200 MHz.
For the HP 8447F option H64 low frequency preamp, set the controls to display 10
through 30 MHz.
NOTE:
The signal into the spectrum analyzer input mixer must be low enough to eliminate
harmonic distortion in the spectrum analyzer.
3
Set the sweep oscillator output level for the following signal levels out of the amplifier
(read on the spectrum analyzer).
4
Note the level of the second harmonic at 800 MHz and the third harmonic at 1200 MHz.
Signal Level
Spectrum Analyzer
Input Attenuation
HP 8447E:
+
8 dBm
50 dB
HP 8447F Power Amp:
+
8 dBm
50 dB
HP 8447F H64 Low freq preamp:
+
10 dBm
50 dB
Содержание 8447D
Страница 5: ...DRAFT 3 Operation and Service Manual ...
Страница 10: ...8 DRAFT Operation and Service Manual General Information Figure 1 Typical Preamplifier Performance Curves ...
Страница 11: ...DRAFT 9 Operation and Service Manual General Information Figure 2 Typical Power Amplifier Performance Curves ...
Страница 13: ...DRAFT 11 Operation and Service Manual General Information ...
Страница 54: ...52 DRAFT Operation and Service Manual Parts and Service Figure 13 Internal Views ...
Страница 58: ...56 DRAFT Operation and Service Manual Parts and Service Figure 18 Amplifier and Power Supply Circuits ...