DRAFT
17
Operation and Service Manual
Performance Tests Procedures
Network analyzer
(0.1-110 MHz)
System with a swept source capable of measur-
ing the amplitude of reflected signals return loss.
Frequency range: 0.1 to 1300 MHz
Resolution: 0.25 dB
Accuracy:
±
6%
HP 8407A
P
Network analyzer
(0.11 -1.4 GHz)
HP 8410A
Phase-magnitude
display
HP 8412A
Harmonic fre-
quency converter
HP 8411A
S-parameter test set
HP 8745A
Universal extension
HP 11604A
Transmission kit
HP 11651A
Accessory kit
HP 11570A
Spectrum analyzer
RF section
Frequency range: 10-1300 MHz
Dynamic range: 60 dB minimum
Scan width: 800 MHz minimum
Spurious responses: <
−
60 dB
Absolute amplitude calibration.
Variable persistence display.
HP 8555A
P
Spectrum analyzer
IF section
HP 8552A
Display section
HP 141T
Noise figure meter
Frequency range: 0.1-1300 MHz
Noise figure range: 0-15 dB
Accuracy:
±
0.5 dB
Excess noise ratio: 6.3
±
0.5 dB
HP 342A
P
VHF noise source
HP 343A
UHF noise source
HP 349A
RMS voltmeter
Voltage range: 1mV to 1V (full range)
Frequency range: 10 Hz to 10 MHz
Meter accuracy:
±5%
HP 3400A
P,A,T
Power meter
Rower range:
−
20 dBm to
+
10 dBm
Frequency range: 10-1300 MHz
Meter accuracy:
±
1%
HP 432A/478A
P
Amplifier
Gain: 24 dB
Noise: <8 dB
Frequency range: 0.1-130 MHz
HP 8447D
P
Table 4
Recommended Test Equipment and Accessories
Instrument Type
Minimum Specifications
Suggested Model
Use
a
Содержание 8447D
Страница 5: ...DRAFT 3 Operation and Service Manual ...
Страница 10: ...8 DRAFT Operation and Service Manual General Information Figure 1 Typical Preamplifier Performance Curves ...
Страница 11: ...DRAFT 9 Operation and Service Manual General Information Figure 2 Typical Power Amplifier Performance Curves ...
Страница 13: ...DRAFT 11 Operation and Service Manual General Information ...
Страница 54: ...52 DRAFT Operation and Service Manual Parts and Service Figure 13 Internal Views ...
Страница 58: ...56 DRAFT Operation and Service Manual Parts and Service Figure 18 Amplifier and Power Supply Circuits ...