I
I
Performing the Operator’s Check
To check the output power
5. Set the power meter calibration factor to the value listed on the power
sensor that corresponds to the frequency you set.
6. Verify that the measured output power meets the instrument specification.
l
If the measured value is less than the
power, turn the front
panel knob until the value measured is at least the
power level.
l
Make sure that the UNLEV annunciator doesn’t light. (If the annunciator
does light, this indicates an unleveled condition.)
7. Repeat steps 3 through 6 for the following setting(s):
2.5
1.5
for HP
instruments only
50 MHz for HP
instruments only
Содержание 83751A
Страница 1: ...User s Guide HP 83751A B and HP 83752AlB Synthesized Sweepers ...
Страница 22: ...Contents 17 3 Typical Phase Noise 10 GHz Carrier 17 12 17 4 Dimensions 17 17 Contents 8 ...
Страница 24: ...1 Performing the Operator s Check ...
Страница 35: ...I 2 Externally Leveling the Sweeper ...
Страница 46: ...3 Generating a Stepped Sweep ...
Страница 48: ...4 Generating a Millimeter Signal ...
Страница 54: ...5 Creating User Flatness Arrays I ...
Страница 59: ......
Страница 63: ...I I 6 Operating a Master Slave Setup ...
Страница 70: ...I 7 Front Rear Panel ...
Страница 100: ...I 8 Instrument State Keys ...
Страница 112: ...I 9 Marker Keys ...
Страница 121: ...10 Modulation Keys ...
Страница 128: ...11 Frequency Keys ...
Страница 141: ...12 Sweep Keys ...
Страница 151: ...13 Power Keys ...
Страница 161: ...14 Entry Keys ...
Страница 162: ...Entry Keys pg48ab Figure 14 1 The Entry Group 14 2 ...
Страница 171: ...I 15 Special Functions ...
Страница 172: ......
Страница 199: ...16 Error Messages ...
Страница 235: ...I 17 Specifications and Options I ...
Страница 244: ...Specifications and Options RF output Source Match 1 7 1 SWR internaLLy Leveled 1 7 1 0 ...
Страница 245: ......
Страница 252: ...Specifications and Options General Weight Net 16 kg 35 lb Shipping 23 kg 49 lb 17 18 ...
Страница 257: ...18 Safety and Regulatory Information ...
Страница 266: ...Index ...