Test 4: Double Pulse Delay
Test Specifications
Equipment Needed
Procedure
Range
20
ns to 999 ms
&solution
3 digits, best case 100 ps
Accuracy
2 5 0
p s
Min. Period
133.4 ns
Digitizing Oscilloscope with Accessories
Counter
Cable, 50 coaxial, BNC
1. Connect HP 8114A to the Scope as shown:
HP 8 4A
TEST
51
FRONT END
A D A P T O R
Connecting
8114A to the Scope
Testing the HP
Содержание 8114A
Страница 14: ......
Страница 24: ......
Страница 118: ......
Страница 162: ......
Страница 168: ......
Страница 174: ......
Страница 176: ......
Страница 178: ......
Страница 180: ......
Страница 182: ......
Страница 183: ...MODEL 8114A INDEX OF MANUAL CHANGE MANUAL FRAME CHANGE ERRATA 1 MP13 3 MPI 4 W6 MPl19 2 September 1997 Page 5 of 6 ...