8114A Service Guide
-200
Execution error
-200
-211
HP-IB message = “Product- or Serial
is wrong.”
Local message = “Product- or Serial Nr. is wrong.”
Description:
The header of the calibration data stream contains
a product number ( Char.) or a serial number (10
Char.). The length of
strings will be checked by
the firmware.
Execution error
HP-IB message = “Security violation”
Local message = “Security violation”
Description:
An attempt was made to switch the display ON, while
security was on.
Trigger ignored
HP-IB message = “Trigger signal ignored”
Local message = “MANUAL key ignored”
Description:
The MANUAL trigger function is not supported by all
modes. The TRG or GET HPIB commands have the
same functionality as the MANUAL key.
- 2 2 1
Settings conflict
HP-IB message = “Width period”
Local message = “Width period”
Description:
-221
Width exceeds period.
Settings conflict
HP-IB message = “Delay period”
Local message = “Delay period”
Description:
-221
Delay exceeds period.
Settings conflict
HP-IB message = “Too many parameters in % of
period/width”
Local message =
many
in of
per/width”
5-24 Troubleshooting