Description
Use this command to program the offset current of the OUTPUT signal.
Note that to set the OUTPUT levels in terms of current, you first have to
execute the
[:SOURce]:HOLD CURRent
command to enable the
:SOURce]:CURRent
subsystem.
The available current range is limited by the combination of:
•
Specified Voltage limits
•
Actual OUTPUT Impedance setting
:OUTPut:IMPedance
•
Actual Expected Load impedance setting
Example
To program the offset current of the output signal:
Command
:CURR[1]:HIGH
Long
[:SOURce]:CURRent[1][:LEVel][:IMMediate]:HIGH
Form
Set & Query
Parameter
Numeric
Parameter suffix
A
with engineering prefixes.
Value coupling
Range coupling
Low level
*RST value
+10 mA (50
Ω
into 50
Ω
)
Specified Limits
10 V Output (from high Z into short): –396 mA to 400 mA typical
3.8 V (from 50
Ω
into short
):
–82 mA to 152 mA typical
Description
Use this command to program the high level current of the OUTPUT
signal. Note that to set the OUTPUT levels in terms of current, you first
:HOLD CURR
:CURR:OFF 50MA
Enable CURRENT subsystem
Set OUTPUT offset to 50 mA
Amplitude = High – Low
Offset =
High – Low
2
Содержание 81101A
Страница 1: ...H HP 81101A 50 MHz Pulse Generator Reference Guide ...
Страница 10: ...10 Safety Information ...
Страница 88: ...88 Programming Reference SCPI Instrument Command List ...
Страница 112: ...112 Specifications Pulse Parameter Definitions ...
Страница 116: ...116 Index ...
Страница 118: ...Hewlett Packard GmbH Copyright 1998 Edition E0998 Printed in Germany September 1998 BAR CODE to go here 81101 91011 ...