37
Programming Reference
Programming the Instrument Trigger Modes
External Width
Set External Width mode using the
:EWIDth[:STATe]
command:
:ARM:EWIDth ON
Switch on EXT WIDTH mode
This command disables the arm-trigger system. The arm-trigger system is
reenabled by switching
OFF EWIDth
mode.
Pulses
Set Pulses mode by setting the
:TRIGger:COUNt
to 1 so that a single
triggered pulse period is generated for every arm event. The trigger
source sets the pulse period:
:TRIGger:COUNt 1
Single pulse period per arm event
:TRIGger:SOURce INTernal1 Pulse period from internal osc.
:DIGital:PATTern OFF
Disable pattern data.
N O T E
The internal PLL (INTernal2)
cannot
be used as
:TRIGger:SOURce
(pulse period source) if it is already being used as
ARM:SOURce
(triggering rate).
Note that in triggered pulses mode the pulse period source is not relevant
because a single pulse is generated for each ARM event.
Burst
Set Burst mode by setting the
:TRIGger:COUNt
to the burst length
required. The trigger source sets the pulse period for the pulses within
the burst.
:TRIGger:COUNt 16
Burst of 16 pulse periods
:TRIGger:SOURce INTernal1 Pulse period from internal osc.
:DIGital:PATTern OFF
Disable pattern data
Pulse period Source
:TRIGger SOURce
internal OSC.
internal PLL
CLK-IN
INTernal[1]
INTernal2
EXTernal2
Содержание 81101A
Страница 1: ...H HP 81101A 50 MHz Pulse Generator Reference Guide ...
Страница 10: ...10 Safety Information ...
Страница 88: ...88 Programming Reference SCPI Instrument Command List ...
Страница 112: ...112 Specifications Pulse Parameter Definitions ...
Страница 116: ...116 Index ...
Страница 118: ...Hewlett Packard GmbH Copyright 1998 Edition E0998 Printed in Germany September 1998 BAR CODE to go here 81101 91011 ...