  |  HoverTech International  

T-BurgManual, Rev.A 

T-Burg

 User Manual

Intended Use and Precautions

INTENDED USE

The HoverMatt® T-Burg™ is intended for all surgical procedures where a 

patient may be placed in Trendelenburg (or Reverse Trendelenburg) up to 

30 degrees, including with use of robotics. Patient transfer / repositioning 

/ boosting can be facilitated before and / or after the procedure where the 

patient’s weight might place staff at risk for injury. The HoverMatt T-Burg is 

designed to securely hold a patient in varying degrees of Trendelenburg, 

up to 30 degrees.

The HoverTech Air Supply inflates the HoverMatt T-Burg to cushion and 

cradle the patient, while air simultaneously escapes from the holes on the 

underside, reducing the force needed to transfer and move a patient by 

80-90% to a stretcher or hospital bed. This can ease the patient transfer 

from pre-op to the O.R. Table, and following the procedure from the O.R. 

Table to PACU.

The HoverMatt T-Burg is a highly breathable, open cell design that 

supports an ideal microclimate allowing  the Mat to stay under a patient 

following surgery.  It is recommended that the HoverCover™ Disposable 

Absorbent Cover (or equivalent) be placed under the patient after surgery 

if the Mat is soiled.

INDICATIONS

•   Patients who will be put into the Trendelenburg (or Reverse 

Trendelenburg) position with or without Lithotomy or Tilt.

•   Patients who are unable to assist in lateral transfer, repositioning and / 

or boosting.

•   Patients whose weight or girth poses a potential health risk for the 

caregiver(s) responsible for repositioning and/or laterally transferring 

said patient.

CONTRAINDICATIONS

•   Patients over 400 lbs (182 kgs). Patients that have unstable thoracic, 

cervical or lumbar fractures should not use the HoverMatt T-Burg 

unless a clinical decision has been made by your facility.

INTENDED CARE SETTINGS

•  Hospitals and Ambulatory Surgery Centers (ASC)

PRECAUTIONS – TRENDELENBURG 

(REVERSE) POSITION (WITH OR WITHOUT TILT)

•   Caregivers must verify that all connecting straps are tightly connected 

to the O.R. table to ensure the HoldFast™ Strip is in contact with the 

O.R. Table.

•  Attach patient safety (body) straps across the patient.

WARNING:

 Do not exceed a 30° angle.

IMPORTANT:

 THE PATIENT’S SKIN MUST MAKE DIRECT CONTACT WITH 

THE HOLDFAST™ SURFACE.

PRECAUTIONS – O.R. TO PACU TRANSFERS

•   Caregivers must verify that all caster brakes have been engaged prior 

to transfer.

•  Use a minimum of three caregivers during air-assisted lateral patient   

      transfers. 
•  Never leave the patient unattended when the device is  inflated.
•   Use this product only for its intended purposes, as described in this 

manual.

•   Only use attachments and/or accessories that are authorized by 

HoverTech International.

•   When transferring to a low air loss bed, set the bed mattress air flow to 

the highest level for a firm transfer surface.

•  Never attempt to transfer a patient on an uninflated HoverMatt T-Burg.

WARNING:

 Prior to moving a patient into a Trendelenburg (or Reverse 

Trendelenburg) position, ensure (i) that the Mat is deflated and (ii) the 

Mat is secured to the O.R. Table per instructions on pg.4, and (iii) that the 

patient is secured to the Mat and O.R. Table itself per hospital guidelines 

and (iv) ensure that instructions are followed and in accordance with 

hospital knowledge, practices and guidelines for placing patients safely in 

Trendelenburg.

PRECAUTIONS – AIR SUPPLY

•   Not for use in the presence of flammable anesthetics or in a hyperbaric 

chamber or oxygen tent.

•  Route the power cord in a manner to ensure freedom from hazard.
•  Avoid blocking the air intakes of the HT-Air® Air Supply.
•   When using the HoverMatt T-Burg in an MRI environment, a 25 ft. 

specialty MRI hose is required (purchased separately).

CAUTION:

 Avoid electric shock. Do not open Air Supply container.

WARNING:

 Reference your Air Supply’s product-specific user manual for 

operating instructions.

Symbol Reference

DECLARATION OF CONFORMITY  

TO MEDICAL DEVICE DIRECTIVE

Air Hose Port 

(1 per side)

Содержание HOVERMATT T-Burg

Страница 1: ...g to Operating Room O R Table 5 Part Identification HT Air Air Supply 6 HT Air Keypad Functions 6 Air200G Air400G Air Supplies 6 Product Specifications Required Accessories 7 Cleaning and Preventive M...

Страница 2: ...mbulatory Surgery Centers ASC PRECAUTIONS TRENDELENBURG REVERSE POSITION WITH OR WITHOUT TILT Caregivers must verify that all connecting straps are tightly connected to the O R table to ensure the Hol...

Страница 3: ...neum Cut out Transfer Handles 2 per side Rail Fasteners 3 per side Foot end Straps 1 per side HoldFast Surface Fastening Strip HoldFast Strip Patient Safety Body Strap 2 parts Head end Part Identifica...

Страница 4: ...foot end of HoverMatt T Burg snap into place 6 Turn on HoverTech Air Supply 7 Push HoverMatt T Burg at an angle either headfirst or feetfirst across the surfaces Once half way across caregiver on the...

Страница 5: ...ctors purchased separately around each arm Ensure that arms are securely positioned against the patient s body per hospital procedures protocols 14 Position table and patient for the procedure 15 Foll...

Страница 6: ...the inflation speed by the number of flashes i e two flashes equals the second inflation speed All of the settings in the ADJUSTABLE range are substantially lower than the HoverMatt and HoverJack set...

Страница 7: ...stem not to function as intended the Mat should be removed from use Returns and Repairs All products being returned to HoverTech International HTI must have a Returned Goods Authorization RGA number i...

Отзывы: