Preparation
23
Condition settings
Standard calibration conditions and measurement conditions are set by default in the product.
Once the warm-up icon has turned off, calibration and measurement can be started
immediately.
Calibration condition settings
The default settings for the calibration conditions and the pages to refer to for the setting
procedures are shown below.
Measurement condition settings
The default settings for the measurement conditions and the pages to refer to for the setting
procedures are shown below.
Measurement
condition
Default setting
Setting procedure page
Span Point
200.0 mg/L
Extraction Time
40 sec
Separation Time
30 sec
Number of Purge
2 time
Calib. Mode
AUTO
Measurement
condition
Default setting
Setting procedure page
Extraction Time
40 sec
Separation Time
30 sec
Fill Cell Time
60 sec
Meas. Limit
300 sec
Drainage Time
30 sec
Number of Purge
2 time
Meas. Mode
AUTO
Measurement Unit
mg/L
Solvent Vol.
8.0 mL
Sample Vol.
16.0 mL
Zero Shift Value
0.0 mg/L
Use Light
ON
Confirm Save
AUTO
Save Memo
OFF
Display Negative
OFF
Display Raw Data
OFF