Setting
73
Fill Cell Time
Use this screen to set the cell filling time in the automatic measurement and automatic
calibration modes.
Fig. 74 Fill Cell Time screen
The following settings are available. The default setting is 60 (sec).
For the buttons and functions, which can be used with this screen appearing, refer to Table 3
(page 17).
Meas. Limit
Use this screen to set the maximum measurement time in measurement and calibration.
If the measured value does not stabilize, measurement ends when the set time (from the start
of measurement or calibration) elapses, and the measured value at that point is displayed.
Fig. 75 Meas. Limit screen
The following settings are available. The default setting is 300 (sec).
For the buttons and functions, which can be used with this screen appearing, refer to Table 3
(page 17).
Setting range
Units
30 to 3600
sec
Setting range
Units
60 to 3600
sec