Setting
76
Solvent Vol.
Use this screen to enter the solvent volume to be used as a coefficient for conversion of the
measurement unit into mg/kg, mg/g, or mg/PC.
Reference
"Conversion of measurement units" (page 112)
Fig. 80 Solvent Vol. screen
The following settings are available. The default setting is 8.0 (mL).
For the buttons and functions, which can be used with this screen appearing, refer to Table 3
(page 17).
Sample Vol.
Use this screen to enter the sample volume to be used as a coefficient for conversion of the
measurement unit into mg/kg, mg/g, or mg/PC.
Reference
"Conversion of measurement units" (page 112)
Fig. 81 Sample Vol. screen
The following settings are available. The default setting is 16.0 (the units depend on the set
[Measurement Unit]).
For the buttons and functions, which can be used with this screen appearing, refer to Table 3
(page 17).
Setting range
Units
1.0 to 1000.0
mL
Setting range
Units
1.0 to 10000.0
The units depend on the set [Measurement Unit] (refer to "Measurement Unit" (page 75)).
kg if [Measurement Unit] is set to mg/kg
g if [Measurement Unit] is set to mg/g
PC if [Measurement Unit] is set to mg/PC