Setting
77
Zero Shift Value
Use this screen to set the shift correction value for zero liquid.
The sum of this set value and the raw measured value is displayed as the measured value.
Reference
"Conversion of measurement units" (page 112)
Fig. 82 Zero Shift Value screen
The following settings are available. The default setting is 0.0 (mg/L).
For the buttons and functions, which can be used with this screen appearing, refer to Table 3
(page 17).
Use Light
Use this screen to set ON or OFF for the LED that illuminates the inside of the extraction tank.
Turning ON the LED makes it easier to check liquid amounts and the boundary surface
through the extraction tank window.
Fig. 83 Use Light screen
The following settings are available. The default setting is "ON".
Tip
Even if [Use Light] is set to "ON", the LED will be turned OFF automatically when the period of the set
[B-Light Off Time] has passed after the last button operation (refer to "B-Light Off Time" (page 85)).
Any button operations turn ON the LED again.
For the buttons and functions, which can be used with this screen appearing, refer to Table 2
(page 16).
Setting range
Units
-100.0 to 100.0
mg/L
Selection
Description
OFF
Turns OFF the LED that illuminates the inside of the extraction tank.
ON
Turns ON the LED that illuminates the inside of the extraction tank.