![horiba LAQUAact D-74 Скачать руководство пользователя страница 69](http://html1.mh-extra.com/html/horiba/laquaact-d-74/laquaact-d-74_instruction-manual_2162256069.webp)
Printing measured values and calibration data
D-74
53
< Inspection data (pH) >
< Calibration data (conductivity) >
Printout format
Description
Inst. model
:
D-74
Instrument model
Inst. SN
:
KL1TSN20
Instrument serial number
Calibration data
Date
:
2013/09/01
Calibration date
Time
:
09
:
00
Calibration time
Inspection before use
Repeatability
:
0.01
Repeatability of calibration data
pH 6.85
:
7.8 mV
Measurement potential at time of
repeatability inspection
:
25.1°C ATC
Temperature value and temperature setting
Printout format
Description
Inst. model
:
D-74
Instrument model
Inst. SN
:
KL1TSN20
Instrument serial number
CELL
:
Cell constant
1.000×100 m
-1
Calibration data
Date
:
2013/09/01
Calibration date
Time
:
09
:
00
Calibration time
140.9 mS/m
:
Calibration value
:
25.0°C ATC
Temperature value and temperature setting
Temp Coef
:
2.00%/°C
Temperature coefficient