Printing measured values and calibration data
50
HORIBA
< Measurement data >
Printout format
Description
Date
:
2013/09/01
Measurement date
Time
:
09
:
00
Measurement time
Channel
:
1
Measurement Channel
pH
:
5.22
Measured value
HOLD
:
AUTO
HOLD status
Hold value: AUTO
Not hold value: INST
Temperature
:
25.5°C ATC
Temperature value and temperature setting
Sample
:
0000
Sample ID number
Inst. model
:
D-74
Instrument model
Inst. SN
:
KL1TSN20
Instrument serial number
Elect. status
:
OK
Electrode status based on calibration result
Offset
:
0.7 mV
Asymmetry potential of calibration data
Sensitivity
Sensitivity of calibration data
pH 4.01 - 6.86
:
98.9%
Date
:
2013/09/01
Measurement date
Time
:
09
:
00
Measurement time
Channel
:
2
Measurement Channel
COND
:
1.121 mS/m
Measured value
HOLD
:
AUTO
HOLD status
Hold value: AUTO
Not hold value: INST
Temperature
:
25.5°C ATC
Temperature value and temperature setting
Sample
:
0000
Sample ID number
Inst. model
:
D-74
Instrument model
Inst. SN
:
KL1TSN20
Instrument serial number
CELL
:
Cell constant
1.000×100 m
-1
Temp Coef
:
2.00%/°C
Temperature coefficient
Содержание LAQUAact D-74
Страница 1: ...Portable pH ORP Conductivity Meter D 74 Instruction Manual CODE GZ0000333608...
Страница 2: ......
Страница 56: ...M E M O 40 HORIBA...
Страница 80: ...M E M O 64 HORIBA...
Страница 88: ...M E M O 72 HORIBA...
Страница 98: ...M E M O 82 HORIBA...
Страница 110: ...M E M O 94 HORIBA...
Страница 111: ...2 Miyanohigashi Kisshoin Minami ku Kyoto 601 8510 Japan http www horiba com...
Страница 112: ......