Displaying the latest calibration and inspection data
46
HORIBA
・
Electrode status based on calibration result
• Displaying the latest inspection data
1. Press the
key to enter the data mode.
2. Press the
keys to select "CHECK" (display inspection data) and then press
the
key.
The repeatability value (the absolute value of the difference between the measured
value and standard solution value) is displayed.
Display
Description
Reference
GOOD
Electrode sensitivity: 93% to 100%
Good condition.
-
CHECK
Electrode sensitivity: 90% to 93%
Urgent measures are not required but attention is
necessary.
P. 66
BAD
Electrode sensitivity: 85% to 90%
Check the electrode immediately.
P. 66
key
key
key