Chapter 4 Maintenance and Troubleshooting
4.2 Troubleshooting
DS-71
59
●
The indicated value does not change/No response.
Cause
Countermeasure
Key operation is
invalid.
Disconnect and reconnect the AC
adapter, and press the ON/OFF key 10
seconds later.
If the same error occurs again, seek
repairs at contact HORIBA sales
representative.
The electrode con-
nector is not
attached correctly.
Turn the power OFF, then turn it back ON
again.
The electrode is
defective.
Replace the electrode.
The pH meter is
defective.
Contact HORIBA sales representative.
●
The details when the AC adapter is disconnected.
Cause
Countermeasure
No lithium battery
is set.
ithium battery
Set the l
(CR2032).
The lithium battery
runs down.
ithium battery
Replace the l
(CR2032).
Date Setting" P.28, and set date and time.
Lithium battery replacement
Lithium battery