Chapter 3 Basic Operation
3.2 Settings
16
HORIBA
4.
If you select MTC, press the
▲
▼
or
key to select
a digit and press the
▲
or
▼
key to change the
value, and then press the ENTER key.
Setting range: 0
°
C to 100
°
C
ATC
In Automatic Temperature Compensation (ATC), the
instrument detects the solution temperature. When the
temperature sensor is connected, the current tempera-
ture of the solutions is displayed automatically. If the
sensor is not connected, the set value for MTC (default:
25
°
C) is displayed.
MTC
In Manual Temperature Compensation (MTC), the tem-
perature sensor of the electrode is not used, and the
solution temperature is entered manually.