Printing measured values and calibration data
D-72
37
< The data saved in internal memory >
< Calibration data (pH, inspection was performed) >
Printout format
Description
Memory Num
:
0001
Data number
Date
:
2013/09/01
Measurement date
Time
:
09
:
00
Measurement time
Channel
:
1
Measurement channel
pH
:
5.22
Measured value
HOLD
:
INST
HOLD status
Hold value: AUTO
Not hold value: INST
Temperature
:
25.0°C MTC
Temperature value and temperature setting
Sample
:
0000
Sample ID number
Inst. model
:
D-72
Instrument model
Inst. SN
:
KL1TSE09
Instrument serial number
Elect. status
:
OK
Electrode status based on calibration result
Printout format
Description
Inst. model
:
D-72
Instrument model
Inst. SN
:
KL1TSE09
Instrument serial number
Elect. status
:
OK
Electrode status based on calibration result
Offset
:
0.7 mV
Asymmetry potential of calibration data
Sensitivity
Sensitivity of calibration data
pH 4.01 - 6.86
:
98.9%
Repeatability
:
0.01
Repeatability of calibration data
Calibration data
Date
:
2013/09/01
Calibration date
Time
:
09
:
00
Calibration time
pH 4.01
:
176.9 mV
Measurement potential
:
25.0°C ATC
Temperature value and temperature setting
pH 6.86
:
7.6 mV
Measurement potential
:
25.1°C ATC
Temperature value and temperature setting
Inspection before use
pH 6.85
:
7.8 mV
Measurement potential at time of
repeatability inspection
:
25.1°C ATC
Temperature value and temperature setting