Deleting calibration data
D-72
35
●
Deleting calibration data
You can delete the calibration data set in the instrument.
1. Press the
key to enter the setting mode.
2. Press the
keys to select the "MAINT" (maintenance setting) and then press
the
key.
3. Press the
keys to select the "CALCLR" (delete calibration data) and then
press the
key.
4. Select "YES" to delete the calibration data, or select "NO" to cancel deleting it.
And then press the
key.
When "YES" is selected, "END" appears after deletes calibration data. Press the
key.
To return to the setting mode, press the
key.
key
key
key
key
Selected "YES"
and press the
key