DO Tip (7544)
Instruction Manual
Thank you for purchasing the HORIBA DO electrode. Read
this manual carefully before using the electrode.
*
"DO" is an abbreviation of Dissolved Oxygen.
Conformable standards
This equipment conforms to the following standards:
Authorised representative in EU and UK
HORIBA Europe GmbH
Hans-Mess-Str.6, D-61440 Oberursel, Germany
HORIBA UK Limited
Kyoto Close, Moulton Park, Northampton, NN3 6FL, UK
Cautions during handling
An extremely thin membrane is stretched across the DO
tip and DO responsive part. Do not allow the membrane
to be in contact with any hard surface and do not exert
undue force on the diaphragm.
When attaching the DO tip to the electrode body, check
to make sure that O-ring is attached. Attach the DO tip
securely to the electrode body.
DO tips that are no longer usable should be disposed of
as industrial waste.
Packaged contents
Specifications
RoHS: EN
IEC
63000
9. Monitoring and control instruments including
industrial monitoring and control instruments
RoHS: BS
EN
IEC
63000
9. Monitoring and control instruments including
industrial monitoring and control instruments
WARNING
Chemical solution
Highly concentrated potassium hydroxide (KOH) is used in the
internal solution of the electrode.
If the internal solution in the electrode comes in contact with your
hands or skin, wash immediately with water. If the internal solution
comes in contact with your eyes, flush immediately with large
amounts of water and seek treatment by a physician. Follow the
instructions on the associated Safety Data Sheet (SDS).
Name
Q'ty
DO tip (7544)
1 pc
Instruction manual
1 copy
Packing box
1 pc
Model
7544
Measurement principles
Membrane galvanic cell
Measurement range
0 mg/L to 20.00 mg/L
Usable temp. range
0°C to 50°C
Materials
PPS
*
, Sn, Ag
Ext. diam. of wetted part
15.6 mm
*
: Wetted material
Storage
The electrode should remain to be connected to the
meter during storage.
To store the electrode without the meter, follow the
below procedure.
1. Remove the electrode form the connector on the meter.
2. Clean the electrode well with tap water, then dry it with a
soft cloth.
3. Remove the DO tip from the holder.
4. Put the short socket on the DO tip.
5. Store the DO tip in a closed container in a cool, dark
location with an oxygen scavenger.
Note
Avoid storing the electrode in hot place or places with high
humidity. Sore the electrode indoors and out of direct
sunlight.
The color of the internal solution may change depending on
storage conditions but does not affect performances of the
product.
Installation procedure
1. Remove the DO tip (7544) form the package.
2. Remove the short socket from the DO tip.
3. Attach the sensor adapter to the holder when measuring
in BOD bottle.
4. Attached the DO tip to the holder.
5. Put the stirrer into the stirrer holder.
6. Attach the stirrer holder to DO tip.
Note
Do not dispose of the short socket that was removed. It is
necessary when storing the electrode.
When attaching the DO tip, make sure that the O-ring is
attached to the DO tip, and screw the DO tip on firmly, until
it stops.
Disposal method
When disposing of the electrode, follow the related laws
and/or regulations of your country/region.
Short socket
DO tip
O-ring
Holder
Sensor adapter
Holder
DO tip
Stirrer holder
Stirrer
Stirrer holder
DO tip