Table 4 

Recommended separation distances between 

  portable and mobile RF communications equipment 

and the device 

The device is intended for use in an electromagnetic environment 
in which radiated RF disturbances are controlled. The user of this 
device can help prevent electromagnetic interference by 
maintaining a minimum distance between portable and mobile RF 
communications equipment (transmitters) and this device as 
recommended below, according to the maximum output power of 
the communications equipment. 

Rated 

maximum 

output 

power of 

transmitte

Separation distance according to 

frequency of transmitter 

150 kHz 

to 80 

MHz 

 

80 MHz 

to 800 

MHz 

 

800 MHz to 2.5 

GHz 

0,01 

N/A 

0.12 

0.23 

0,1 

N/A 

0.38 

0.73 

N/A 

1.2 

2.3 

10 

N/A 

3.8 

7.3 

37 

 

Содержание Newborn & Mom Pulse OX 594020

Страница 1: ......

Страница 2: ...d reproduced or translated into another language without the prior written consent We reserve the right to improve and amend it at any time without prior notice This device complies with Part 15 of th...

Страница 3: ...e when ambient temperature exceeds 98 6 F 37 C Injury may occur due to overheating of the sensor Discomfort may occur if the oximeter is used on the same finger for an extended period of time especial...

Страница 4: ...the sensor in the correct position The patient s finger should remain still during measurement Do not put wet finger directly into sensor Do not let anything block the emitting light from device Vigor...

Страница 5: ...eater or strong sunshine etc may cause inaccurate measurement Please put an opaque cover on the sensor or change the patient s location If the first reading shows unstable waveform the measurement may...

Страница 6: ...tion 2 1 5 Features 2 3 2 Battery Installation 3 3 Operation 4 3 1 Direct measurement 4 6 3 2 Alarm and alarm silence 7 3 3 Menu Screen 8 9 3 4 Hopkins Y Probe Sensor 10 11 3 5 Proper application of H...

Страница 7: ...Pulse Ox 6 1 Oximeter Maintenance 20 21 6 2 Cleaning and Disinfecting Instruction 22 6 3 Replacement Parts 22 23 7 Troubleshooting 24 Appendix 25 A Key of Symbols 25 B SpO2 Common Knowledge 25 30 C El...

Страница 8: ...User Manual for the Hopkins Newborn Mom Pulse Ox 1 Overview 1 1 Appearance Figure 1 Front Rear View 1...

Страница 9: ...patient s finger It is applicable for spot checking SpO2 and pulse rate of adult and pediatric patients in homes and clinics 1 4 Structure and Conformation It consists of main unit and photoelectric...

Страница 10: ...Battery Installation Figure 2 Battery Installation 1 Refer to Figure 2 insert two AAA size batteries into the battery compartment properly 2 Replace the cover Please make sure that the batteries are c...

Страница 11: ...ower on automatically in 2 seconds and start to display software version number 4 The user can read the values and view the waveform on the display screen as seen in figures 4A 4F below Shift paramete...

Страница 12: ...parameter display between PR and PI is under the condition of the screen tilting to a side as shown in 4B and 4E After 20 seconds without operation the PI will automatically shift to PR as shown in 4C...

Страница 13: ...he Hopkins Newborn Mom Pulse Ox Screen Description SpO2 SpO2 symbol 99 SpO2 value PR Pulse rate icon 65 Pulse rate value Pulse beat symbol Pulse intensity histogram PI Perfusion index icon 1 4 Perfusi...

Страница 14: ...on the screen will flash The detailed information refers to chapter 4 To silence the alarm or re set the pulse ox 1 Press Display Key to mute If this alarm event continues the pulse ox will resume al...

Страница 15: ...o enter the menu screen Menu screen description 1 SpO2 alm Lo SpO2 alarm Lower limit The user can modify the value of 85 99 the step is 1 the default is 90 2 PR alm Hi Pulse Rate alarm Upper limit The...

Страница 16: ...isplay screen 6 Restore default Restore default setting Refer to Figure 5 for each default value On setup menu screen 1 Press the Display Key to choose the setting item 2 Press and hold the Display Ke...

Страница 17: ...gure 6 Probe Connection Important Note When the Hopkins Y Probe Sensor is connected properly the built in finger clip sensor will be disabled The measurement is detected from the external SpO2 probe 2...

Страница 18: ...ls in the patient The recommended application site for infant and pediatric patients is on the finger foot or hand The recommended application for adult patients is on the finger Using the Hopkins Y P...

Страница 19: ...sensor centered in the middle of it will minimize any effects from ambient light sources Other wraps can be used to secure lead wires to minimize the movement of the sensor itself IMPORTANT LED Senso...

Страница 20: ...x Free Belt and tighten gently 5 Apply a Hopkins Y Probe wrap on top to prevent ambient light interference A secure wrap with the probe sensor centered in the middle of it will minimize any effects fr...

Страница 21: ...in water or a mild soap solution 3 To disinfect wipe the sensor and patient contact surfaces with disinfecting solution Isopropyl alcohol is recommended If using Sanizide follow the manufacturer s re...

Страница 22: ...t 3 Important Batteries are not to be stored in the oximeter for an extended period of time and battery failure due to prolonged storage in the oximeter is not covered under the 3 year product warrant...

Страница 23: ...l Incubator Continuous use of fingertip SpO2 sensor may result in discomfort or pain so we suggest inspecting the monitoring site for skin integrity every 2 hours for an adult patient every 1 hour for...

Страница 24: ...sensor wavelength Red light 663 nanometers Infrared light 890 nanometers Maximal optical output power less than 1 5mW maximum average B SpO2 measurement Measuring range 70 100 Measuring accuracy 70 99...

Страница 25: ...Display Range 0 2 20 E Audible visual alarm function When measuring if SpO2 value or pulse rate value exceeds the preset alarm limit the device will alarm automatically and the value which exceeds li...

Страница 26: ...between the SpO2 value measured in the condition of indoor natural light and that of darkroom is less than 1 When using the Hopkins Wrap ensure the probe is centered in the wrap to prevent interferenc...

Страница 27: ...6 Maintenance Replacement Parts 6 1 Oximeter Maintenance The service life not a warranty of this device is 5 years In order to ensure its long service life please adhere to the use of maintenance Plea...

Страница 28: ...he factory prior to sale there is no need to calibrate the device during use However if it is necessary to verify its accuracy routinely the user can obtain verification by means of SpO2 simulator or...

Страница 29: ...soft cloth wet with clean water and let air dry or wipe it dry You may use Sanizide to clean the unit Caution Do not sterilize by irradiation steam or ethylene oxide Do not use the sensor if it is da...

Страница 30: ...User Manual for the Hopkins Newborn Mom Pulse Ox Hopkins Y probe wraps 592012BG is a bag of 100 23...

Страница 31: ...r correctly inside and try again 2 Keep the patient calm Cannot turn the device on 1 The batteries are drained or almost drained 2 The batteries are not inserted properly 3 The device is malfunctionin...

Страница 32: ...g See User Manual SpO2 The pulse oxygen saturation PR Pulse rate beats per minute Pulse rate icon Low battery voltage B SpO2 Common Knowledge 1 Meaning of SpO2 SpO2 is the saturation percentage of oxy...

Страница 33: ...al with its density or concentration When the light with certain wavelength emits on human tissue the measured intensity of light after absorption reflecting and attenuation in tissue can reflect the...

Страница 34: ...gical parameter to analyze a patient s respiration and blood oxygenation levels Some examples where SpO2 monitoring is important are monitoring patients with serious respiratory disease patients under...

Страница 35: ...g lamps or direct sunlight Insufficient coverage of the y probe can interfere with readings Make sure to have the Y Probe centered in the middle of the Hopkins Wrap Vascular dyes or externally used pr...

Страница 36: ...User Manual for the Hopkins Newborn Mom Pulse Ox Blood vessel contraction caused by peripheral vessel hyperkinesia or body temperature decreasing 29...

Страница 37: ...Hypoxemia disease functional lack of HbO2 Pigmentation or abnormal oxyhemoglobin level Abnormal oxyhemoglobin variation Methemoglobin disease Sulfhemoglobinemia or arterial occlusion exists near sens...

Страница 38: ...l for the Hopkins Newborn Mom Pulse Ox C Guidance and Electromagnetic Interference Data The following tables Table 1 to 4 provide the specific information regarding this device s compliance to IEC 606...

Страница 39: ...idance RF emissions CISPR 11 Group 1 This device uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic...

Страница 40: ...uch an electromagnetic environment Immunity test IEC 60601 test level Compli ance level Electromagnetic environment guidance Electrostati c discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV conta...

Страница 41: ...4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 s N A N A Power frequency 50 60Hz magnetic field IEC61000 4 8 3 A m 3 A m...

Страница 42: ...ce level Electromagnetic environment guidance Conduct ed RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 Vrms 150 kHz to 80 MHz 3 V m 80 N A 3 V m Portable and mobile RF communications equipment should...

Страница 43: ...ects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast c...

Страница 44: ...device can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and this device as recommended below accordi...

Страница 45: ...pplicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separa...

Отзывы: