Surge
IEC
61000-4-5
±1 kV line
(s) to line(s)
±2 kV line(s)
to earth
N/A
N/A
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC
61000-4-11
<5 %
U
T
(>95 % dip
in
U
T
)
for 0.5 cycle
40 %
U
T
(60 % dip in
U
T
)
for 5 cycles
70 %
U
T
(30 % dip in
U
T
)
for 25 cycles
<5 %
U
T
(>95 % dip
in
U
T
)
for 5 s
N/A
N/A
Power
frequency
(50/60Hz)
magnetic
field
IEC61000-4-
8
3 A/m
3 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or
hospital
environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test
level.