PPLICATION OF THE IDROMED
®
5
PS
2.1.
F
IELDS OF
A
PPLICATION
Condition:
Used for the treatment of the following conditions:
•
Hyperhidrosis
2.2.
C
ONDITIONS OF
A
PPLICATION
idromed
®
5 PS
has been designed for operation in dry rooms with an ambient temperature from
10°C to 45°C.
Protect the unit against chemical vapours and never operate it in explosive areas.
2.3.
P
REPARATION OF THE
T
REATMENT
The
idromed
®
5 PS
is operated with the power supply (IDROMED AC/AC ADAPTER, Typ: FM
6299M). After the following preparations your unit is ready for use:
a) For the best results and for your own security, the unit should only be used with
original accessories from the manufacturer.
b) For treatment you should separate the two parts of the case. To do this, open the case
completely, put it on a plain surface and separate the 2 parts carefully in opposite directions
(see arrows on the hinges on the case).
c) Fill the 2 treatment tubs with tap water to a depth of approx. 3-4 cm (finger and toe joints
should be just covered). Please see also the warnings on page 10, point 4.3.
d) Connect one of the cables to one of the plate electrodes and to the
(+)
-connection on the
front of the
idromed
®
5 PS
.
e) Connect the 2nd cable of the electrode with the other electrode and with the
(-)
-connection on
the front of the
idromed
®
5 PS
.
f)
Place the electrodes 1 in each tub and cover the electrodes with the distance grids.
g) Connect the plug of the power supply to the rear of the
idromed
®
5 PS
and plug the power
supply to a mains socket (220-230 V / 50 Hz). The green control lamp on the power pack is
now alight.