ROBLEMS AND HOW TO SOLVE THEM
PROBLEM
SOLUTION
4.1. The green lamp on
the mains plug
(IDROMED ADAPTER)
does not work.
In brightly lit rooms, it can be difficult to see whether the lamp is on or
not. Shade the lamp with your hand to check, whether it is working
properly.
The wall socket may be defective. Please connect the unit to another
wall socket or check whether the wall socket in question is working
properly.
4.2. The unit shows on
the display:
Malfunction!
Switch off the machine!
At a failure malfunction the unit’s output current stops automatically and
the unit displays a malfunction message. Switch off the unit and switch it
on again. If the error lamp is still alight, please contact Customer
Services.
4.3. Current intensity is
adequate.
Check the connections of the plug and electrodes and the function of
the unit.
Please check if the water you are using has been treated with a water
softener. By using softened water, the conductivity is reduced. In this
case please use untreated tap water or use non-carbonated (still)
mineral or table water.
For the treatment of the armpits, check if the sponges are still moist
enough, as in the description under point 2.4.2.
If current intensity is not adequate despite the above mentioned points,
please contact the Customer Service.
4.4. The display shows
nothing.
Check if the lamp is lit on the plug unit (see point 4.1.). Check the
functioning of the socket.
Press the on/off switch several times (turning knob).
If you have further problems please send the unit to the manufacturer
for repair.