1
O
PERATING
I
NSTRUCTIONS
idromed
®
5 PS
Art.-No. 110001
Art.-No. 110003
Before putting the unit into operation and starting treatment, read this operating manual carefully!
#110001/3 / S0408 e
Страница 1: ...1 OPERATING INSTRUCTIONS idromed 5 PS Art No 110001 Art No 110003 Before putting the unit into operation and starting treatment read this operating manual carefully 110001 3 S0408 e ...
Страница 2: ...arge treatment tubs art no 995130 2 x distance grids for large treatment tubs art no 995131 Content 1 Important information on idromed 5 PS 3 1 1 Symbols and warning on the unit 4 1 2 Electromagnetical compatibility 4 2 Application of the idromed 5 PS 5 2 1 Fields of Application 5 2 2 Conditions of Application 5 2 3 Preparation of the Treatment 5 2 4 Treatment 6 2 4 1 Treatment of hyperhidrosis on...
Страница 3: ...d during treatment All jewellery on hands and feet must be removed before starting treatment Any damage to skin minor injuries scratches etc on the palms of the hands soles of the feet cuticles or at the armpits must be coated with vaseline or a fatty cream before treatment as current permeability will be higher than normal on those spots Please do only cover these spots Current intensity is set a...
Страница 4: ...cology Unit contains electronic parts and may not be disposed in the household rubbish 1 2 ELECTROMAGNETICAL COMPATIBILITY idromed 5 PS uses high frequency energy exclusively for its internal function Therefore its HF emission is very low and it is unlikely that surrounding electronic devices are disturbed HF emissions acc to CISPR 11 group 1 idromed 5 PS is determined for the use in all facilitie...
Страница 5: ...e manufacturer b For treatment you should separate the two parts of the case To do this open the case completely put it on a plain surface and separate the 2 parts carefully in opposite directions see arrows on the hinges on the case c Fill the 2 treatment tubs with tap water to a depth of approx 3 4 cm finger and toe joints should be just covered Please see also the warnings on page 10 point 4 3 ...
Страница 6: ...reatment of the feet or of hands and feet 25 mA For the treatment of the armpits 5 mA If the current intensity selected is too high you may experience an unpleasant tingling sensation in the extremities treated during the session If this occurs a lower current intensity has to be selected otherwise there is a danger of burns 2 4 1 TREATMENT OF HYPERHIDROSIS ON THE HANDS AND OR FEET Self test Maxim...
Страница 7: ...would break the circuit by taking one hand out of the tubs While adjusting the current hands or feet may be taken out of the treatment tubs and while the unit is still running without any danger Please dry your hands before regulating the current at the unit Regulate the current by turning the knob right or left and close the circuit again by putting again your hands in the treatment tubs Repeat t...
Страница 8: ...ect the electrodes with the positive and the negative pole of the unit Self test Maximum armpits 5mA Values hands 15mA feet 25mA Switch on the unit while pushing down the turning knob Pay attention that the turning knob is on switch setting The unit carries out a self test This action takes some seconds Set the therapy current Place your hands feet or armpit sponges 2mA Adjust the required current...
Страница 9: ...ns the remaining treatment time At the end of the treatment remove the sponge electrodes from the armpits Turn the knob to the switch setting and switch off the unit by pushing down the knob Always check the position of the turning knob Position immediately after switching on the unit and the self test completes 3 IMPORTANT TIPS AND INFORMATION To simplify therapy the current intensity in mA for y...
Страница 10: ...please contact Customer Services 4 3 Current intensity is adequate Check the connections of the plug and electrodes and the function of the unit Please check if the water you are using has been treated with a water softener By using softened water the conductivity is reduced In this case please use untreated tap water or use non carbonated still mineral or table water For the treatment of the armp...
Страница 11: ...lly the general Terms of Sales and Delivery of the company Dr K Hönle Medizintechnik GmbH shall apply These shall be submitted to the operator upon the signing of the contract at the very latest The warranty for the idromed 5 PS is 24 months Warranty and liability claims are excluded in case of personal injury or damage of property if these are attributable to one or more of the following causes T...
Страница 12: ...l be alight Once the unit has been switched off the lamp will be out Protection class II all insulated MDD classification IIa Protection against intrusion of liquids IPX1 CE Mark Output wave form Manufacturer A L T Lichttherapietechnik GmbH Thura Mark 10 D 06780 Zörbig Exclusive sales Dr K Hönle Medizintechnik GmbH Viktor Frankl Str 20 D 86916 Kaufering Tel 0049 8191 30555 0 Fax 0049 8191 30555 50...