Page 54
ST 800 Series HART/DE Option User’s Manual
Revision 5.0
Deleting offline configuration
Using this feature you can delete an offline configuration template.
To delete an offline configuration, perform the following steps.
12.
On the FDC homepage, tap Offline Configuration > Select.
The
Offline Configurations
page appears.
13.
Select the required offline configuration template from the list.
14.
Tap
Options
>
Delete
. A warning message appears.
15.
Tap
Yes
to delete the offline configuration template.
Downloading an offline configuration
Using this feature, you can download the offline configuration when the device is online.
To download an offline configuration, perform the following steps.
16.
On the FDC homepage, tap
Offline Configuration > Select
.
The
Offline Configurations
page appears.
17.
Select the required offline configuration template from the list.
18.
Tap
Options
>
Download
.
The
Offline – Select Variables
page appears with the all the variables.
Note
: By default, all the variables selected in FDM will appear as selected and non-editable
variables appear in grey color.
19.
Select the required variable. In case you select a dependent variable, then variables on which it
is dependent on will also be selected and the following warning appears.
20.
Tap
OK
to return to the offline wizard.
21.
Tap
Next
.
The
Offline – Review and Send
page appears with the list of selected variables.
22.
Tap
Send
and the process to send the variables to the device starts. Once the downloading is
complete, the following page appears.