7. Operation
7.2. Transmitter PV display
48
SmartLine Wireless Temperature and Universal I/O User's Manual
Revision 1
Transmitter display
OneWireless User
Interface display
Definition
What to do
‘U’ Status
Excess Span
Calibration
The calibrated upper
and lower trim has
produced a span that is
greater than 5% of the
characterized span of
the transmitter.
Clear Calibration
Or
Set Factory Calibration
Or
Check the applied trim points
and re-attempt lower and
upper (trim) calibration.
‘U’ Status
Excess Calibrated
Range
The selected
calibration points used
for upper and lower
trim are outside the
characterized range of
the transmitter.
Check that the upper and
lower trim points are both
within the characterized range
of the transmitter and re-
attempt upper and lower (trim)
calibration.
blank
Calibration Cleared
Indicates that both the
upper and lower trim
points as well as the
zero offset has been
cleared. The calibration
source is none.
Calibration the zero offset
Or
Calibrate using the lower and
upper trim points.
Содержание SmartLine STIW400
Страница 92: ...80 SmartLine Wireless Temperature and Universal I O User s Manual Revision 1 B5 Control Drawing ...
Страница 93: ...Revision 1 SmartLine Wireless Temperature and Universal I O User s Manual 81 ...
Страница 94: ...82 SmartLine Wireless Temperature and Universal I O User s Manual Revision 1 ...
Страница 95: ...Revision 1 SmartLine Wireless Temperature and Universal I O User s Manual 83 ...
Страница 96: ...84 SmartLine Wireless Temperature and Universal I O User s Manual Revision 1 ...
Страница 97: ...Revision 1 SmartLine Wireless Temperature and Universal I O User s Manual 85 ...
Страница 98: ...86 SmartLine Wireless Temperature and Universal I O User s Manual Revision 1 ...
Страница 99: ...Revision 1 SmartLine Wireless Temperature and Universal I O User s Manual 87 ...
Страница 100: ...88 SmartLine Wireless Temperature and Universal I O User s Manual Revision 1 ...
Страница 101: ...Revision 1 SmartLine Wireless Temperature and Universal I O User s Manual 89 ...