Mounting
The Extractive Module XT is prepared for DIN-rail mounting. An appropriate mounting rail
`
is delivered with the instrument, which allows mounting the Extractive Module XT and the
Satellite XT side by side.
Mount this rail to a wall with the appropriate fasteners and slide both modules onto it.
Plug in male and female connector of Satellite XT and Extractive Module XT.
Module Connections
sample tubing.
sensor connection between Extractive Module XT and
Satellite XT.
Data link and power supply.
9.6
Tubing Connections
Tubing material is PTFE or PFA with an outer diameter of 1/4” and an inner diameter of
3/16”. Length and inner diameter of the sample line will affect the response time. Length of
the sample lines should be kept as short as possible. With a sample line ID of 3/16”
(4.7 mm) and a length of 10 m, the delay in response does not exceed 30 seconds.
Sample Line Inlet
The fitting for connecting the sample line inlet is located on top of the housing and marked
“Gas Inlet”. Improper installation of the sample inlet tube may result in dilution or even
complete loss of the sample.
Exhaust Line Outlet
The fitting for connection of the exhaust line outlet is located on top of the instrument and
marked “Outlet”. Since hazardous gas concentrations may still be present at the exhaust
line outlet, it is recommended to connect the exhaust line to an exhaust duct.
! Note:
Do not apply power until the system is ready for start-up.