![Honeywell Satellite XT 9602-0200 4-20 mA Скачать руководство пользователя страница 43](http://html.mh-extra.com/html/honeywell/satellite-xt-9602-0200-4-20-ma/satellite-xt-9602-0200-4-20-ma_operation-manual_151884043.webp)
Satellite XT 4-20 mA Versions
Auto Selftest
The instrument's self-diagnostics provides an on-line
preventive sensor selftest performed automatically
every 24 hours, which may be enabled or disabled.
Default setting is enabled.
Use the <set> key to enter. The display will show the
actual setting. The function Auto Selftest is not avail-
able with instruments using oxygen sensors or catalytic
sensors.
Disabled
Allows the user to turn off the Auto Selftest.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
Enabled
Allows the user to turn on the Auto Selftest.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
Security
Exiting the Monitoring Mode and entering the Mainte-
nance Mode, may be password protected, to prevent
the instrument from manipulations by unauthorized per-
sonnel. Ex factory the password protection is disabled.
Use the <set> key to enter. The instruments shows the
actual setting.
Password On
Allows the user to enable the password protection.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit.
Password Off
Allows the user to disable the password protection.
Actuate the cursor keys <up> or <down> to change.
Actuate the <set> key to confirm or <esc> to exit..
s e l e c t :
4
o f 1 3
C O N F I G U R A T I O N
AU TO S E L F T E S T
s e l e c t :
1
o f
2
A U T O S E L F T E S T
D I S A B L ED
s e l e c t :
2
o f
2
A U T O S E L F T E S T
ENA B L ED
s e l e c t :
5
o f 1 3
C O N F I G U R A T I O N
S ECUR I T Y
s e l e c t :
1
o f
2
S E C U R I T Y
P A S SWORD ON
s e l e c t :
2
o f
2
S E C U R I T Y
P A S SWORD OF F
6-9
Configuration - Section 6