Satellite XT 4-20 mA Versions
Fault - Extractive !
Only for instruments with Extractive Module.
This message indicates that there is a problem with the
Extractive Module. Check power supply, sample tubing,
sampletubing connections, and sample line filter. Perform any
corrective action required.
Output current is 1.2 mA.
If the message is still shown, please contact our service
department for further instructions.
Fault - Pyrolyzer !
Only for instruments with Pyrolyzer Module.
This message indicates that there is a problem with the
Pyrolyzer Module.
Output current is 1.0 mA.
Please contact our service department for further
instructions.
- - -